Doctors Review System, OB-Windows/OB-View software release numbers 4.1 through 4.1.3, Digisonics, Inc., 3701...
Digisonics, Inc began a recall of Doctors Review System, OB-Windows/OB-View software release numbers 4.1 through 4.1.3, Digisonics, Inc., 3701 Kirby Drive, Houston, TX 77098 OB-Windows provides... on . The reason FDA records is "Study would on occasion have fetal heart rate and/or cervical length on the report that was not entered nor did it belong to that patient. Software displayed incorrect data for fetal heart rate and cervical length in OB...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2113-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Digisonics, Inc
- Product
- Doctors Review System, OB-Windows/OB-View software release numbers 4.1 through 4.1.3, Digisonics, Inc., 3701 Kirby Drive, Houston, TX 77098 OB-Windows provides professional reporting and analysis of fetal growth throughout the gestational period, which can help evaluate clinical situations. OB-View provides image management for the user to capture, view, read, and edit images.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Study would on occasion have fetal heart rate and/or cervical length on the report that was not entered nor did it belong to that patient. Software displayed incorrect data for fetal heart rate and cervical length in OB ultrasound reports.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- The firm, Digisonics Inc, notified the affected customers of software release via e-mail on January 24, 2007 and follow-up e-mail on February 7, 2007. The e-mails described the product, problem and the actions taken. The firm was able to trap the error and after this report the developer was able to determine the cause and a new executable was sent and installed at the customers site. If you have any question, please call 800-940-3240 or e-mail: [email protected].
- Quantity in commerce
- 128
- Distribution
- Worldwide distribution: USA (nationwide) and country of Canada.
- Product codes and lots
- Software release 4.1.3