Z-2154-2011

Enzymatic Creatinine (ECREA) Flex(R) reagent cartridges, K1270 for the Dimension Vista(R) 500 System The...

Siemens Healthcare Diagnostics Inc. / initiated 2010-06-10 / Terminated / root cause: design

Siemens Healthcare Diagnostics, Inc. began a recall of Enzymatic Creatinine (ECREA) Flex(R) reagent cartridges, K1270 for the Dimension Vista(R) 500 System The ECREA method is an in vitro diagnostic test for the... on . The reason FDA records is "Results may be falsely depressed up to 0.4 mg/dL or falsely elevated up to 0.6 mg/dL across the assay range". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2154-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Enzymatic Creatinine (ECREA) Flex(R) reagent cartridges, K1270 for the Dimension Vista(R) 500 System The ECREA method is an in vitro diagnostic test for the quantitative measurement of creatinine in human serum, plasma and urine on the Dimension Vista(R) System. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for other urine analytes.
Product code
JFY Enzymatic Method, Creatinine
Reason for recall
Results may be falsely depressed up to 0.4 mg/dL or falsely elevated up to 0.6 mg/dL across the assay range.
Root cause tag
design
FDA root cause
Device Design
Action
Siemens sent an Urgent Field Safety Notice letter dated June 2010 to all Dimension Vista(R) ECREA customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised to: Discontinue use of ECREA method and transition to the Dimension(R) Jaffe creatinine (CREA) method. To notify anyone to whom they may have distributed the product. Complete and return the Effectiveness Check Questionnaire attached to the letter by fax to (302) 631-8467. For questions regarding this recall call (302) 631-6299.
Quantity in commerce
1,399
Distribution
Nationwide Distribution - including AL, AZ, CA, CO, DE, FL, GA, ID, IL, IA, MD, MA, MI, MT, NH, NJ, NY, NC, OH, OR, PA, SC, TX, VA, WA, and WI.
Product codes and lots
Catalog number K1270 - lot numbers 09224AA, 09286AA, 09350AC, 10011AF, 10047AA, 10089AC, 10089AD, 10131AC and 10131AD.