Z-2180-2011
Digisonics, DigiView and ERS software release 3.6.4.4 through 3.6.5.2+, 2008 Digisonics, Inc., 3701 Kirby...
Digisonics, Inc / initiated 2009-01-20 / Terminated / root cause: design
Digisonics, Inc began a recall of Digisonics, DigiView and ERS software release 3.6.4.4 through 3.6.5.2+, 2008 Digisonics, Inc., 3701 Kirby Drive, Houston TX 77098, 800-940-3240. DigiView is a... on . The reason FDA records is "Data from one patient's study had the potential for being put into another patient's report". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2180-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Digisonics, Inc
- Product
- Digisonics, DigiView and ERS software release 3.6.4.4 through 3.6.5.2+, 2008 Digisonics, Inc., 3701 Kirby Drive, Houston TX 77098, 800-940-3240. DigiView is a digital image management and reporting system with an associated clinical database. It provides capture, review, and storage of images together with report creation and data basing.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Data from one patient's study had the potential for being put into another patient's report.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Digisonics notified end users via e-mail January 20, 2009 and January 21, 2009, that Data from one patient's study was being put into another patient's report. The problem occurred when a System bug was found in DigiView Release 3.6.4.4, and higher versions in which under certain circumstances (specific workflow) a failure to switch to a patient's images occurs. Two software patch versions will be made available to sites on or after February 5, 2009, to correct the reported problem. If you have any further questions, please contact Digisonics Customer Support at (800 ) 940-3240.
- Quantity in commerce
- 70 units
- Distribution
- Worldwide Distribution -- USA including States of : OH, NC, LA, TX, CA, MO, NY, MD, MA, DC, IA, AZ, PA, MI, NC, FL, IN, SC, ID, MN, AR, OK, IL, CT, and countries of Canada, Lebanon, and New Zealand.
- Product codes and lots
- DigiView Rel. 3.6.4.4, 3.6.4.6, and 3.6.5.2.