Z-2166-2011

DigiView software release 3.4.4.4 through 3.5.5.2. , 2008 Digisonics, Inc, 800-940-3240. Digiview is a...

Digisonics, Inc / initiated 2008-11-04 / Terminated / root cause: design

Digisonics, Inc began a recall of DigiView software release 3.4.4.4 through 3.5.5.2. , 2008 Digisonics, Inc, 800-940-3240. Digiview is a digital image management and reporting system with an... on . The reason FDA records is "Failure of software to update some measurements when tracings or points are amended after measurement calculations have been performed". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2166-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Digisonics, Inc
Product
DigiView software release 3.4.4.4 through 3.5.5.2. , 2008 Digisonics, Inc, 800-940-3240. Digiview is a digital image management and reporting system with an associated clinical database. It provides capture, review and storage of images together with report creation and data basing.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Failure of software to update some measurements when tracings or points are amended after measurement calculations have been performed.
Root cause tag
design
FDA root cause
Device Design
Action
Digisonics, Inc., notified consignees by telephone and e-mail on 10/17/2008, informing them that a problem had been found in the Save Linked Images functionality of DigiView Release 3.6 (and all releases which start with 3.6 such as 3.6.1.). The problem is the failure of the software to update some measurements when tracings or points are amended after measurement calculations have been performed. A utility program, called Linked Image Measurement Verify was being supplied to all sites as a patch to the DigiView program. Release 3.6.6 was to be provided to all customers having previous 3.6 version. Problem was fixed in version 3.6.6 and 3.7.1. If you have any questions, please contact Digisonics Customer Support at (800) 940-3240.
Quantity in commerce
70 units total
Distribution
Worldwide Distribution -- USA including States of : TX, OH, NC, MA, NY, IL, FL, OK, MO, ID, AZ, AR, MI CA, MD, and country of Canada. *** Due to the age of this recall, district is recommending no audit checks***
Product codes and lots
DigiView software release numbers 3.6.4.4, 3.6.4.6, and 3.6.5.2