Boston Scientific, Atlantis SR Pro 2, 40 MHz Coronary Imaging Catheter, 3.6F (1.18mm) x 135 cm, USE BY...
Boston Scientific Corporation began a recall of Boston Scientific, Atlantis SR Pro 2, 40 MHz Coronary Imaging Catheter, 3.6F (1.18mm) x 135 cm, USE BY 2011-05, REF 39014, UPN H749390140, Sterile R Sterilized... on . The reason FDA records is "Catheter tip detachment that is not associated with elongation of the catheter typically caused by excessive pulling forces. EXPANDED: Boston Scientific is initiating a recall of all iCross Coronary Imaging Catheters...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2419-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Boston Scientific, Atlantis SR Pro 2, 40 MHz Coronary Imaging Catheter, 3.6F (1.18mm) x 135 cm, USE BY 2011-05, REF 39014, UPN H749390140, Sterile R Sterilized using irradiation, Made in USA 47900 Bayside Parkway, Fremont, CA 94538 The intended use as stated in Japan Shonin 21200BZY00484000: This product is an imaging catheter with built-in ultrasonic transducer that sends ultrasound waves inside the blood vessel to perform B-mode ultrasonagraphy. The indications for use as stated in the 510(k) and DFU 90606867: The Atlantis coronary catheters are intended for ultrasound examination of coronary intravascular pathology only. Intravascular Ultrasound imaging is indicated in patients who are candidates for transluminal coronary interventional procedures.
- Product code
- DQO Catheter, Intravascular, Diagnostic
- Reason for recall
- Catheter tip detachment that is not associated with elongation of the catheter typically caused by excessive pulling forces. EXPANDED: Boston Scientific is initiating a recall of all iCross Coronary Imaging Catheters because they have received and confirmed 8 occurrences of catheter tip detachments due to embrittlement.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Boston Scientific " Important Customer Notification" letters dated March 14, 2011 were hand delivered to customers by Boston Scientific sales representatives starting on March 28th, 2011. The sales representatives verbally reviewed information. A tracking log will be maintained document completion of the customer notifications. The letter described the product, problem and possible root cause as well as one corrective action, a change in sterilization method from Gamma to E-beam implementation. Boston Scientific sent an Urgent Medical Device Recall Removal - Immediate Action Required letter dated May 26, 2011, to all affected customers. Customers were instructed to Identify and discontinue use of any affected product within their inventory and to segregate it immediately and return it to Boston Scientific
- Quantity in commerce
- 30,669
- Distribution
- Worldwide Distribution - USA (nationwide) and Puerto Rico and the countries of Japan, Trinidad, Tobago, and American Virgin Islands.
- Product codes and lots
- All lots are impacted