Z-2106-2011
Clearview iFOB, Clearview iFOB Complete- Immunochemical Fecal Occult Blood Cassette Test Product Part Number...
Abbott Rapid Dx North America, LLC / initiated 2011-03-21 / Terminated
Alere North Americal began a recall of Clearview iFOB, Clearview iFOB Complete- Immunochemical Fecal Occult Blood Cassette Test Product Part Number: 8111KCV (30 Test Kit ), Intended Use... on . The reason FDA records is "Potential for false negatives test results and negative/faint positive controls". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2106-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Rapid Dx North America, LLC
- Product
- Clearview iFOB, Clearview iFOB Complete- Immunochemical Fecal Occult Blood Cassette Test Product Part Number: 8111KCV (30 Test Kit ), Intended Use: Immunochemical device intended for the qualitative detection if fecal occult blood in human feces by laboratories or physician offices
- Product code
- KHE Reagent, Occult Blood
- Reason for recall
- Potential for false negatives test results and negative/faint positive controls
- FDA root cause
- Other
- Action
- Alere sent a Voluntary Product Recall letter dated March 21, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to indicate zero on the attached Recall Reply Form, sign, and fax to 1-877-877-0147, if they have no remaining product. For any remaining product customers were instructed to discontinue use and indicate the quantity of kits in the attached Recall Reply Form so that the firm can arrange for them to receive credit. Customers were instructed to destroy any affected units at their site. For any further distribution of product, customers will need to be contacted. Physicians and hospitals were advised to retest patients with negative results. Questions relating to this Voluntary Product Recall call 1-877-597-9586.
- Quantity in commerce
- 6249 kits
- Distribution
- Nationwide Distribution
- Product codes and lots
- Lot Numbers: 0616210 0626710 0818710 1021010