Z-2193-2011

Digisonics, Digiview and ERS software releases 3.6.44, 3.6.4.6, and 3.6.5, Digisonics, Inc., 3701 Kirby...

Digisonics, Inc / initiated 2009-02-06 / Terminated

Digisonics, Inc began a recall of Digisonics, Digiview and ERS software releases 3.6.44, 3.6.4.6, and 3.6.5, Digisonics, Inc., 3701 Kirby Drive, Houston, TX 77098, Tel: (800) 940-3240, Fax... on . The reason FDA records is "Left ventricle volume was inaccurate when the volumes were measured from images having different scales". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2193-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Digisonics, Inc
Product
Digisonics, Digiview and ERS software releases 3.6.44, 3.6.4.6, and 3.6.5, Digisonics, Inc., 3701 Kirby Drive, Houston, TX 77098, Tel: (800) 940-3240, Fax: (713) 529-7999 Software used to make on-screen (2-D, M-Mode and Doppler) measurements of left ventricle, right ventricle, etc.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Left ventricle volume was inaccurate when the volumes were measured from images having different scales.
FDA root cause
Other
Action
The firm, Digisonics, notified consignees/customers via e-mail entitled "Distribution of Notification of Biplane Error" dated February 6, 2009. The email described the product, problem and actions to be taken. The customers were instructed to immediately contact Digisonics if their site calculates biplane volumes-a utility will be provided which will run on their database and identify all studies in which biplane volumes are calculated. The customers were also instructed to share the following important information with all users of DigiView at their site. The email stated -until the software patch has been installed, biplane volumes should be calculated only from pairs of images having the same depth scale. Replacement DigiView/ERS software that prevents further occurrence of this problem will be provided if it appears their site could be impacted. The Distribution of New Software Release was sent on March 6, 2009. If you have any questions, please call Digisonics Customer Support at 800-344-4266
Quantity in commerce
74 units
Distribution
Worldwide distribution: USA (Nationwide) and countries of: Canada, Lebanon and New Zealand.
Product codes and lots
DigiView software release numbers 3.6.44, 3.6.4.6, and 3.6.5