PERPOS Driver; Catalog # 6112 Interventional Spine, Inc. A cannulated, driver to be used over a K-wire to...
Interventional Spine Inc began a recall of PERPOS Driver; Catalog # 6112 Interventional Spine, Inc. A cannulated, driver to be used over a K-wire to engage the head of the 4.5mm BONE-LOK PLS Implant... on . The reason FDA records is "This recall has been initiated due to drivers that may contain a burr in the distal tip preventing full seating of the driver onto the head of the facet screw, resulting in difficulty screwing the facet screw into the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0047-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Interventional Spine, Inc.
- Product
- PERPOS Driver; Catalog # 6112 Interventional Spine, Inc. A cannulated, driver to be used over a K-wire to engage the head of the 4.5mm BONE-LOK PLS Implant (LSW-45-3040) and drive it into the bone. The PERPOS Driver (61 12) is sold as either a component of the Single Use PERPOS PLS System (9045-01, 9045-02), a component of the Reusable BONE-LOK PLS System (9039-01, 9039-02), or as a stand-alone tool.
- Product code
- HXX Screwdriver
- Reason for recall
- This recall has been initiated due to drivers that may contain a burr in the distal tip preventing full seating of the driver onto the head of the facet screw, resulting in difficulty screwing the facet screw into the bone. Use of this product may result in a delay of the surgical procedure
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Interventional Spine, Inc. sent an "Urgent Recall" letter dated March 15, 2011 to all affected customers. The letter included; affected product, reason for recall and recommendations. Customers were asked to immediately examine inventory, quarantine affected product and to notify International Spine, Inc if any product remain. For information on this recall contact the firm at (949) 472-0006.
- Quantity in commerce
- 12 units
- Distribution
- Word wide Distribution: (USA) Nationwide including the states of CO, CT, FL, KY, LA, MA, NC, NY, OK, OR, TN, TX, VA, and WI; and the countries of Italy, Spain, Turkey, and the U.K.
- Product codes and lots
- DRIVER CATALOG NO. 6112; 050507A, 061509A, 071808A, 112507A, 042709E