BardVentralex" Hernia Patch Medium Circle with Strap Diameter 6.4cm /2.5",\ Catalog #0010302, lot number...
Davol, Inc., Sub. C. R. Bard, Inc. began a recall of BardVentralex" Hernia Patch Medium Circle with Strap Diameter 6.4cm /2.5",\ Catalog #0010302, lot number HUTL1283. The BardVentralex" Hernia Patch is a... on . The reason FDA records is "Mislabeled: Bard Medium Ventralex Hernia Patch 6.4cm /2.5" packaged in a pre-printed carton identified for a Bard Ventralex Small Circle 4.3cm /1.7"". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2124-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Davol Inc., Sub. C.R. Bard, Inc.
- Product
- BardVentralex" Hernia Patch Medium Circle with Strap Diameter 6.4cm /2.5",\ Catalog #0010302, lot number HUTL1283. The BardVentralex" Hernia Patch is a self-expanding, non-absorbable, sterile prosthesis. The device classification name is "Mesh, Surgical, Polymeric". It is intended for use in all forms of hernia repair requiring reinforcement with a non-absorbable support material
- Product code
- FTL Mesh, Surgical, Polymeric
- Reason for recall
- Mislabeled: Bard Medium Ventralex Hernia Patch 6.4cm /2.5" packaged in a pre-printed carton identified for a Bard Ventralex Small Circle 4.3cm /1.7"
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- Davol sent an Urgent: Medical Device Recall letter on April 11, 2011, via FedeX Priority Overnight to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to examine their inventory and remove and return the recalled lot. Distributors were requested to notify their accounts. For questions contact the Davol Customer Service Department at 1-800-556-6275 or C. R. Bard's Medical Services & Support Department at 1-800-562-0027.
- Quantity in commerce
- 146 units
- Distribution
- Nationwide Distribution - USA including AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, MA, MD, MI, MO, MT, NC, NJ, NV, NY, OH, OR, PA, RI, SC, TN, TX, VA, WA, and WV.
- Product codes and lots
- Lot number HUTL1283
FIRST ARTICLE / WEEKLY