Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-3188-2011

Orthovita(R) Cortoss(TM) Delivery Gun

DSM Biomedical, Inc. / initiated 2011-03-31 / Terminated / root cause: packaging

Orthovita, Inc. began a recall of Orthovita(R) Cortoss(TM) Delivery Gun Product Usage: Used for the application of Cortoss(TM) Bone Augmentation Material on . The reason FDA records is "Recalling firm has discovered through internal testing that the Cortoss(TM) Delivery Gun may abrade its packaging material under certain transportation and handling conditions. This abrasion may cause small holes or...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-3188-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
DSM Biomedical, Inc.
Product
Orthovita(R) Cortoss(TM) Delivery Gun Product Usage: Used for the application of Cortoss(TM) Bone Augmentation Material
Product code
KIH Dispenser, Cement
Reason for recall
Recalling firm has discovered through internal testing that the Cortoss(TM) Delivery Gun may abrade its packaging material under certain transportation and handling conditions. This abrasion may cause small holes or tears in the packaging material, which may not be immediately visible to the user. Although t the Cortoss(TM) Delivery Gun is provided in double pouches and there have been no report
Root cause tag
packaging
FDA root cause
Packaging
Action
Orthovita sent an "URGENT FIELD SAFETY NOTICE" letter dated March 31, 2011 to all customers. The letter described the product, problem and actions to be taken. The customers were instructed to Identify, quarantine and discontinue use of the affected product. The letter included an acknowledgement form to be completed and returned as per instructions.. The letter states that the recalling firm will contact customers to arrange for the return or destruction of any affected product. A pre-paid shipping label is provided for product return. The letter also states that the recalling firm will provide replacement product at no cost. For any questions regarding this notice call 484-323-8860.
Quantity in commerce
3174
Distribution
Product was distributed within the US to medical facilities in AL, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, NE, NH, NJ, NY, NC, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, and WI. Government accounts include Bethesda Naval Center, 8901 Rockville Pike, Bethesda, MD; Lackland Air Force Base, 2200 Berquist Drive, Suite 1, Lackland AFB, TX; Hunter Holmes McGuire VA Hospital 1201 Broad Rock Blvd., Richmond, VA; Madigan Army Medical Center, Madigan Army Medical Center, Tacoma, WA; Ralph H. Johnson VA Medical Center, 109 Bee Street, Charleston, SC; VA Medical Center, 1481 W. 10th St., Indianapolis, IN; VA Hospital of Palo Alto, 3801 Miranda Ave., Palo Alto, CA; VA Loma Linda Healthcare System, 11201 Benton Street, Loma Linda, CA; and VA Medical Center/NY Harbor Healthcare, NY, NY.
Product codes and lots
Part Number 2110-0008: Lot numbers affected are A812057, A902032, A902063, A904005, A905009, A905039, A906004, A910016, A910021, A912029, A912039, and A1001020. No other lot numbers are affected.