Peloris II Rapid Tissue Processor; Leica Biosystems Melbourne Pty Ltd, Mount Waverly, 3149 VIC, Australia...

Leica Microsystems, Inc. / initiated 2011-04-11 / Terminated / root cause: manufacturing process

Leica Microsystems, Inc. began a recall of Peloris II Rapid Tissue Processor; Leica Biosystems Melbourne Pty Ltd, Mount Waverly, 3149 VIC, Australia; catalog number 26.0005. This tissue processor... on . The reason FDA records is "The processor potentially has a faulty impellor which could lead to reagent carry-over from one processing step to another". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2597-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Leica Microsystems, Inc.
Product
Peloris II Rapid Tissue Processor; Leica Biosystems Melbourne Pty Ltd, Mount Waverly, 3149 VIC, Australia; catalog number 26.0005. This tissue processor prepares tissue samples for sectioning by transforming fixed samples into wax embedded samples. This is achieved by exposing the tissue samples to a sequence of reagents in the processing retorts.
Product code
IEO Processor, Tissue, Automated
Reason for recall
The processor potentially has a faulty impellor which could lead to reagent carry-over from one processing step to another,
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Leica Microsystems sent a MEDICAL DEVICE CORRECTION letter dated April 11, 2011, to the affected customers via first class mail. The letter identified the product, the problem, and the action needed to be taken by the customer. The Customers were instructed to discontinue use of the device until their local Leica Microsystems representative contact them to arrange the replacement of the faulty impellor. The accounts were asked to complete the attached Acknowledgement Form, listing the serial numbers of the units at their facility and fax it back to Leica at 1-847-607-3147. Additional questions should be addressed to the firm's office of Regulatory Affairs and Quality Assurance.
Quantity in commerce
22 units
Distribution
Worldwide Distribution -- USA (nationwide) and the countries of Australia, Canada, Germany Japan, Portugal and United Kingdom.
Product codes and lots
catalog number 26.0005, serial numbers 0265050B, 0265051B, 0265053B, 0265054B, 0265055B, 0265056B, 0265057B, 0265058B, 0265059B, 0265060B, 0265061B, 0265063B, 0265064B, 0265067B, 0265068B, 0265071B and 0265072B.