Siemens brand ACUSON S2000, Model 10434730, ABVS ultrasound systems; Affected sub-components Numbers: (1)...
Siemens Medical Solutions USA, Inc. began a recall of Siemens brand ACUSON S2000, Model 10434730, ABVS ultrasound systems; Affected sub-components Numbers: (1) 10434968, Boom Welding (for S/Ns 250157, 250162... on . The reason FDA records is "The weld connecting the ABVS arm to the system base has the potential to fail, causing the arm to fall off of the unit". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2150-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens brand ACUSON S2000, Model 10434730, ABVS ultrasound systems; Affected sub-components Numbers: (1) 10434968, Boom Welding (for S/Ns 250157, 250162, 250163, 250164, 250166, 250167 and 250169); (2) 10434969, Arm Welding (for S/N's 250171, 250173, 250177); Product is manufactured and distributed by Siemens Medical Solutions USA, Inc., Concord, CA The S2000 Automated Breast Volume Scanner (ABVS) Ultrasound System for imaging the breast allows volumetric acquisition of the breast image. It provides the operator with the option of using a controlled user independent automated scan motion. The resulting 3D display information enables comprehensive image review by the medical professional. In particular, the coronal plane is scanned, which is not easily accessible with hand-held ultrasound transducers.
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- The weld connecting the ABVS arm to the system base has the potential to fail, causing the arm to fall off of the unit.
- Root cause tag
- manufacturing process
- FDA root cause
- Process change control
- Action
- Siemens Medical Solutions USA, Inc. initiated the recall via phone calls and e-mail on March 28, 2011, to all affected customers, informing them of the affected product and providing instructions on the recall. Three units S/N 250171, 250173, and 250177 were to be removed and shipped back to Siemens. Siemens Service personnel were to be dispatched to the customer sites to perform validated stress test on units S/N 250157, 250162, 250163, 250164, 250166, 250167, and 250169 at customer sites. For questions regarding this recall call 650-694-5993.
- Quantity in commerce
- 10
- Distribution
- Worldwide Distribution - USA, Portugal, Germany, Russia, Estonia, Brazil, France, and Saudi Arabia.
- Product codes and lots
- Serial Numbers Affected: 250171, 250173, 250177, 250157, 250162, 250163, 250164, 250166, 250167 and 250169