syngo Dynamics version 9.0, Siemens Medical Solutions, USA, Inc., Ann Arbor, MI. The syngo Dynamics is a...
Siemens Medical Solutions, USA, Inc began a recall of syngo Dynamics version 9.0, Siemens Medical Solutions, USA, Inc., Ann Arbor, MI. The syngo Dynamics is a picture archiving and communications system intended... on . The reason FDA records is "If one or more graphs are in a table on the report, and if any of those graphs are deleted, it may lead to reports failing to save and/or marking the study as read might leave the study in an unread state or might mark...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2293-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- syngo Dynamics version 9.0, Siemens Medical Solutions, USA, Inc., Ann Arbor, MI. The syngo Dynamics is a picture archiving and communications system intended for acceptance, transfer, display, storage, archiving and manipulation of digital medical images, including quantification and report generation. It is not intended to be used for the reading of mammography images.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- If one or more graphs are in a table on the report, and if any of those graphs are deleted, it may lead to reports failing to save and/or marking the study as read might leave the study in an unread state or might mark the study as read but leave the report with a "Preliminary" watermark displayed.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm, SIEMENS, sent a "Customer Safety Advisory Notice" letter dated March 24, 2011 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to place graphs in sections and not tables in the report; avoid deleting graphs from the reports in run-time and include this safety advisory notice in their operator's manual where it should remain until the service patch is applied. SIEMENS will be issuing a service patch to fix the defect. The letter states the patch will be available in the next few weeks. If you have any questions, call +1 (734) 205-2400.
- Quantity in commerce
- 29 devices
- Distribution
- Nationwide distribution: USA including states of: CA, CT, FL, ID, MD, MI, MN, NJ, NY and PA.
- Product codes and lots
- 10091602, 10091604