Software versions Eos software versions: 1.2.4, 1.2.5, 1.2.6,1.2.7,1.2.8, 1.2.9,1.2.10, .2.11,1.2.12, 1.2.13...
Eos Imaging began a recall of Software versions Eos software versions: 1.2.4, 1.2.5, 1.2.6,1.2.7,1.2.8, 1.2.9,1.2.10, .2.11,1.2.12, 1.2.13, 1.3.2,1.3.3, 1.3.4, 1.4.0, 1.4.1, 1.4.2. sterEOS... on . The reason FDA records is "Print true size, print requests from the EOS acquisition station or the sterEOS image review station may, in certain cases, be affected by a different enlargement factor during printing. T". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2692-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- EOS imaging
- Product
- Software versions Eos software versions: 1.2.4, 1.2.5, 1.2.6,1.2.7,1.2.8, 1.2.9,1.2.10, .2.11,1.2.12, 1.2.13, 1.3.2,1.3.3, 1.3.4, 1.4.0, 1.4.1, 1.4.2. sterEOS software versions: 1.2.3.3788, 1.2.4.4286, 1.3.2.3373, 1.3.3.3671, 1.3.4.3740, 1.3.5.3915, 1.3.6.4272,1.4.1.4813,1.4.2.4850. EOS system is a digital radiography system used in general radiographic examinations. sterEOS workstation is a medical image review station for acceptance, transfer, display, storage, and digital processing of 2D X-ray images.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Print true size, print requests from the EOS acquisition station or the sterEOS image review station may, in certain cases, be affected by a different enlargement factor during printing. T
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- EOS imaging sent a "FIELD SAFETY NOTICE" letter dated April 13, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to wait for the printing queue to empty before sending a second "print true queue" request with different print settings. Customers should then click the button shown on the instruction sheet to view the print queue. EOS imaging will provide a software update to correct all affected systems at no cost to the customers. An EOS representative will contact customers to schedule a time to upgrade the software. Questions can be answered by calling 33 1 55 25 63 90.
- Quantity in commerce
- 5 EOS Systems, 10 sterEOS workstations
- Distribution
- Worldwide Distribution--USA (nationwide) including the states of CA, IA, MO, VA, and WI, and countries of Canada and Europe.
- Product codes and lots
- Software versions Eos software versions: 1.2.4, 1.2.5, 1.2.6,1.2.7,1.2.8, 1.2.9,1.2.10, .2.11,1.2.12, 1.2.13, 1.3.2,1.3.3, 1.3.4, 1.4.0, 1.4.1, 1.4.2. sterEOS software versions: 1.2.3.3788, 1.2.4.4286, 1.3.2.3373, 1.3.3.3671, 1.3.4.3740, 1.3.5.3915, 1.3.6.4272,1.4.1.4813,1.4.2.4850.