Philips Diagnost 94, Diagnost 96, Diagnost 97 Universal X-Ray System The Diagnost is a multifunctional...
Philips Healthcare Inc. began a recall of Philips Diagnost 94, Diagnost 96, Diagnost 97 Universal X-Ray System The Diagnost is a multifunctional, universal X~ray system with remote controlled... on . The reason FDA records is "Updated Instructions For Use: During an examination where the patient is positioned on a -rotatable- chair with knees pointed to the rear of the system, the patients lower legs or ankles may be injured/broken". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2765-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Diagnost 94, Diagnost 96, Diagnost 97 Universal X-Ray System The Diagnost is a multifunctional, universal X~ray system with remote controlled facilities. Routine examinations that can be performed with the Diagnost include: Gastro-Intestinal examination, Urograpy, Bronchography, Arthrography, Hysterography, Myelography, Lymphography, Skeletal radiography, Tomography and Venography/phlebography. Special procedures that require subtracted and non subtracted digital imaging include: Angiography, Peripheral angiography (bolus chasing) and Vascular and non~vascular interventional procedures
- Product code
- JAA System, X-Ray, Fluoroscopic, Image-Intensified
- Reason for recall
- Updated Instructions For Use: During an examination where the patient is positioned on a -rotatable- chair with knees pointed to the rear of the system, the patients lower legs or ankles may be injured/broken
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Philips Healthcare notified accounts by letter on April 18, 2011. The letter advised users of the problem and provided updated instructions for proper placement of the patient during the examination. An Updated Instruction For Use was provided. For further information and support customers were instructed to contact the Philips representative.
- Quantity in commerce
- 52 units
- Distribution
- Nationwide Distribution
- Product codes and lots
- All serial numbers