Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2241-2011

NeuViz 16 Multi-Slice CT Scanner System is a computed Tomography X-ray system, and has a Part number...

Neusoft Medical Systems Co., Ltd. / initiated 2011-02-09 / Terminated

Philips And Neusoft Medical Systems Co., Ltd. began a recall of NeuViz 16 Multi-Slice CT Scanner System is a computed Tomography X-ray system, and has a Part number 989605858501. The device is labeled in part... on . The reason FDA records is "Dimension measurement displays on a combined image were found inaccurate after the clinician merged multiple image series into one image". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2241-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Neusoft Medical Systems Co., Ltd.
Product
NeuViz 16 Multi-Slice CT Scanner System is a computed Tomography X-ray system, and has a Part number 989605858501. The device is labeled in part: "***Neusoft***NeuViz16 MULTI-SLICE CT SCANNER SYSTEM***MANUFACTURED FOR NEUSOFT MEDICAL SYSTEMS CO., LTD***MANUFACTURED BY PHILIPS AND NEUSOFT MEDICAL SYSTEMS CO., LTD***ADDRESS: NEUSOFT PARK, HUNNAN INDUSTRIAL AREA, SHENYANG***CHINA***".
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Dimension measurement displays on a combined image were found inaccurate after the clinician merged multiple image series into one image.
FDA root cause
Other
Action
Neusoft Medical Systems sent an Urgent Device Correction Notification letter dated February 9, 2011, via certified mail to their distributor NEUISYS. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised to discontinue use of combined images. Instead, use MPR Batch for diagnostic reviews until the planned upgrade is provided to the product. Neusoft Medical System and Philips Medical System released the mandatory Field Change Order (FCO) by March 31, 2011. All nine affected systems will be updated with aforementioned kits in six months after the release. Consignees were instructed to contact Service Support Department by e-mail to [email protected]
Quantity in commerce
9 units distributed in US
Distribution
Nationwide Distribution including CA, GA, LA, NC, OH, TN and TX.
Product codes and lots
Serial Numbers:   N16E090002, N16E090003, N16E090004, N16E090007, N16E090008, N16E090010, N16E100015, N16E100017, and N16E100023.