Z-2184-2011

IGFBP-3 ELISA Reagent Kit; Part Number" DSL-10-6600; Lot Numbers: 992743, 992287, 991916, 991530, 991432...

Siemens Healthcare Diagnostics Inc. / initiated 2011-03-30 / Terminated / root cause: manufacturing process

Beckman Coulter Inc. began a recall of IGFBP-3 ELISA Reagent Kit; Part Number" DSL-10-6600; Lot Numbers: 992743, 992287, 991916, 991530, 991432, 990220 & 890830 This assay is intended for in vitro... on . The reason FDA records is "The recall was was initiated because patient results with the IGFBP-3 ELISA kit may be elevated by 70% compared to stated expected ranges in the Instructions for Use (IFU)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2184-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
IGFBP-3 ELISA Reagent Kit; Part Number" DSL-10-6600; Lot Numbers: 992743, 992287, 991916, 991530, 991432, 990220 & 890830 This assay is intended for in vitro diagnostic use.
Product code
CFL Radioimmunoassay, Human Growth Hormone
Reason for recall
The recall was was initiated because patient results with the IGFBP-3 ELISA kit may be elevated by 70% compared to stated expected ranges in the Instructions for Use (IFU).
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Beckman Coulter sent a Product Corrective Action (PCA) letter with attached customer response form on March 30, 2011 to all customers who have the affected product, IGFBP- 3 Enzyme-Linked immunosorbent Assay ( ELISA) kit .This kit provides materials for the quantitative measurement of IGFBP-3 in serum. The letter informed the customers of the reason for recall, the products affected and actions to be taken. Consignees were asked to share this information with their laboratory staff and retain the notification as part of their laboratory Quality System documentation. If they forwarded any of the affected product listed above to another laboratory, they were asked to provide a copy of the letter to them. Beckman further asked consignees to complete and return the enclosed Response Form within 10 days so they are assured consignees have received this important communication. Questions were directed to the Beckman Coulter Customer Support Center. - Website: http://www.beckmancoulter.com/customersupport/support - Phone, call 1-800-854-3633 in the United States and Canada - Outside the United States and Canada contact local Beckman Coulter representative.
Quantity in commerce
1,212 units
Distribution
Worldwide -- USA, Australia, Austria, Belgium, Brazil, Canada, China, Czech Republic, France, Germany, Greece, Hong Kong, India, Italy, Korea, Republic of, Lebanon, Luxembourg, Morocco, Netherlands, New Zealand, Puerto Rico, Syrian Arab Republic, Taiwan, Thailand, Turkey & United Kingdom
Product codes and lots
Lot Numbers: 992743, 992287, 991916, 991530, 991432, 990220 & 890830