Z-2181-2011

Varian Medical System TrueBeam system for stereotactic radiosurgery and radiotherapy. Model number H19

Varian Medical Systems, Inc. / initiated 2011-04-05 / Terminated / root cause: manufacturing process

Varian Medical Systems, Inc. Oncology Systems began a recall of Varian Medical System TrueBeam system for stereotactic radiosurgery and radiotherapy. Model number H19 on . The reason FDA records is "Position sensor failure mode may result in an inaccurate position calculation. 1. The video returned by one of the two cameras inside the Spectra is all white or all black, and the Spectra stops tracking. Or 2. Certain...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2181-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Varian Medical Systems, Inc.
Product
Varian Medical System TrueBeam system for stereotactic radiosurgery and radiotherapy. Model number H19;
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Position sensor failure mode may result in an inaccurate position calculation. 1. The video returned by one of the two cameras inside the Spectra is all white or all black, and the Spectra stops tracking. Or 2. Certain video intensities are not available on one of the two cameras. The image has abnormally abrupt transitions from dark to light portions, without the normal shades of gray.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Varian Urgent Medical Device Correction, Urgent Field Safety Notice letters were sent to all affected facilities by certified mail on April 5, 2011.
Quantity in commerce
20
Distribution
Nationwide distribution.
Product codes and lots
Model number H19; Code numbers: H191011, H191019, H191027, H191033, H191012, H191021, H191028, H191037, H191013, H191023, H191030, H191043, H191016, H191025, H191031, H191045, H191018, H191026, H191032, H191047.  Affected NDA Serial numbers: P7-03112, P7-03174, P7-03221.