Varian Medical System TrueBeam system for stereotactic radiosurgery and radiotherapy. Model number H19
Varian Medical Systems, Inc. Oncology Systems began a recall of Varian Medical System TrueBeam system for stereotactic radiosurgery and radiotherapy. Model number H19 on . The reason FDA records is "Position sensor failure mode may result in an inaccurate position calculation. 1. The video returned by one of the two cameras inside the Spectra is all white or all black, and the Spectra stops tracking. Or 2. Certain...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2181-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Varian Medical Systems, Inc.
- Product
- Varian Medical System TrueBeam system for stereotactic radiosurgery and radiotherapy. Model number H19;
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Position sensor failure mode may result in an inaccurate position calculation. 1. The video returned by one of the two cameras inside the Spectra is all white or all black, and the Spectra stops tracking. Or 2. Certain video intensities are not available on one of the two cameras. The image has abnormally abrupt transitions from dark to light portions, without the normal shades of gray.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Varian Urgent Medical Device Correction, Urgent Field Safety Notice letters were sent to all affected facilities by certified mail on April 5, 2011.
- Quantity in commerce
- 20
- Distribution
- Nationwide distribution.
- Product codes and lots
- Model number H19; Code numbers: H191011, H191019, H191027, H191033, H191012, H191021, H191028, H191037, H191013, H191023, H191030, H191043, H191016, H191025, H191031, H191045, H191018, H191026, H191032, H191047. Affected NDA Serial numbers: P7-03112, P7-03174, P7-03221.