Z-2292-2011

WaveLight FS200 Femtosecond Laser System Alcon Product Code 8065990714 WaveLight Model 1025 Software Version...

WAVELIGHT GmbH / initiated 2011-04-18 / Terminated / root cause: software

Wavelight AG began a recall of WaveLight FS200 Femtosecond Laser System Alcon Product Code 8065990714 WaveLight Model 1025 Software Version: Version Blue, Service Pack 2. The WaveLight FS... on . The reason FDA records is "The software calculating the patterns were found to has a rounding error for certain parameter ranges. These rounding errors could result in an incomplete cut in the donor or patient tissue". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2292-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
WAVELIGHT GmbH
Product
WaveLight FS200 Femtosecond Laser System Alcon Product Code 8065990714 WaveLight Model 1025 Software Version: Version Blue, Service Pack 2. The WaveLight FS 200 Laser System is an ophthalmic surgical laser indicated for use in refractive surgery for LASIK flap cutting, lamellar and perforating keratoplasty, and intrastromal surgery for implants.
Product code
HNO Keratome, Ac-Powered
Reason for recall
The software calculating the patterns were found to has a rounding error for certain parameter ranges. These rounding errors could result in an incomplete cut in the donor or patient tissue.
Root cause tag
software
FDA root cause
Other
Action
WaveLight AG sent an URGENT MEDICAL DEVICE CORRECTION letter April 18, 2011 to the affected customers. The letter identified the product, problem and the action to be taken by the customer. The letter advised the customers to discontinue keratoplasty cuts with the WaveLight FS200 until the new version can be installed. Alcon will contact you to schedule a visit of a service engineer at your site, to disable the keratoplasty function in your device, until a new software version can be installed to fix this problem. If you have experienced a product problem or adverse event related to this medical device correction, please contact Alcon Medical Safety at 1-800-561-6466, option 3. Customers were requested to complete the attached reply form to acknowledge receipt of this notification and fax it to ( 817) 916-9087.
Quantity in commerce
55 units
Distribution
Worldwide Distribution - USA including states of TX, KS, NC, CA, FL, OH, SC, MD, VA and NY and countries of Switzerland, Germany, Greece, United Kingdom, Ireland, Spain, Italy, Australia, India, Malaysia, France, Finland, Saudi Arabia, Czech Republic, Poland, China and South Korea and Iran.
Product codes and lots
Alcon Product Code 8065990714 WaveLight Model 1025 Software Version: Version Blue, Service Pack 2