Z-2690-2011
LiNA Loop Endoscopic Monopolar Loop for LSH (Laparoscopic Supracervical Hysterectomy). Large loop, 8mm...
LiNA Medical Inc / initiated 2009-12-01 / Terminated / root cause: labeling and UDI
LINA Medical ApS began a recall of LiNA Loop Endoscopic Monopolar Loop for LSH (Laparoscopic Supracervical Hysterectomy). Large loop, 8mm connector, 1 X REF: EL-200-8 and 10 X REF: EL-200-8... on . The reason FDA records is "Potential for the LiNA Loop to break during operating procedure". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2690-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- LiNA Medical Inc
- Product
- LiNA Loop Endoscopic Monopolar Loop for LSH (Laparoscopic Supracervical Hysterectomy). Large loop, 8mm connector, 1 X REF: EL-200-8 and 10 X REF: EL-200-8, Manufactured by LiNA Medical ApS, Surgery Division.
- Product code
- KNF Coagulator-Cutter, Endoscopic, Unipolar (And Accessories)
- Reason for recall
- Potential for the LiNA Loop to break during operating procedure.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- Lina Medical Aps notified customers by newsletter in December 2009 followed by a letter to distributors including warnings added to instructions for use in January 2010.
- Quantity in commerce
- 1,950 pieces
- Distribution
- GA, MN, PA, WA, TN, MA, IN, Ontario Canada
- Product codes and lots
- Batch 9435, 9422, 9404, 9462, 9451, 9452, 9442, 9371, 9343, 9131, 9471, 7282, 8284, 8285, 8411, 8395, and 8442