Z-2542-2011

Bio-modular reverse shoulder humeral tray with lock ring, 44 mm standard, sterile, single use, REF 11-113680...

Zimmer GmbH / initiated 2011-04-22 / Terminated / root cause: manufacturing process

Biomet, Inc. began a recall of Bio-modular reverse shoulder humeral tray with lock ring, 44 mm standard, sterile, single use, REF 11-113680, Biomet Orthopedics, Warsaw, IN. Biomet... on . The reason FDA records is "The firm is initiating this recall following an investigation which identified the rare possibility that certain units of Comprehensive Reverse Shoulder Humeral Trays may contain a locking ring that is incorrectly...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2542-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Bio-modular reverse shoulder humeral tray with lock ring, 44 mm standard, sterile, single use, REF 11-113680, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder products are indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary. The Comprehensive Reverse Shoulder is indicated for primary, fracture, or revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency.
Product code
KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
Reason for recall
The firm is initiating this recall following an investigation which identified the rare possibility that certain units of Comprehensive Reverse Shoulder Humeral Trays may contain a locking ring that is incorrectly assembled.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
The firm, Biomet Orthopedics, sent an "URGENT MEDICAL DEVICE RECALL" letter dated April 22, 2011 to their customers. The letter describes the product, problem, and actions to be taken. The customers were instructed to immediately locate the recalled product; discontinue use; return the product back to Biomet; confirm receipt of the notice by calling 800-348-9500 extension 1368 or 1325, and complete and return the attached form via fax to: 574-372-1683 or mail to: Biomet, 56 East Bell Drive, Warsaw, IN 46580. Customers with questions regarding this recall should call 800-348-9500 extension 1368 or 1325.
Quantity in commerce
1,575 units all products
Distribution
Worldwide distribution: USA (nationwide) including: Puerto Rico and countries including: Australia, Canada, Chile, Columbia, Costa Rica, Korea and Netherlands.
Product codes and lots
M070320