Z-2426-2011

Ringloc + Acetabular Shell limited hole finned 56 mm Size 24 porous coat, for uncemented use, sterile, REF...

Zimmer GmbH / initiated 2011-04-14 / Terminated / root cause: manufacturing process

Biomet, Inc. began a recall of Ringloc + Acetabular Shell limited hole finned 56 mm Size 24 porous coat, for uncemented use, sterile, REF 16-104156, Biomet Orthopedics, Warsaw, IN. Intended... on . The reason FDA records is "The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2426-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Ringloc + Acetabular Shell limited hole finned 56 mm Size 24 porous coat, for uncemented use, sterile, REF 16-104156, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plasma Spray (PPS) Ringloc Acetabular System is intended to replace the natural acetabulum damaged by disease, trauma or revision of previous arthroplasty. Indications For Use: 1 . Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. 2. Rheumatoid arthritis 3. Correction of functional deformity 4. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques. 5. Revision procedures where other treatment or devices have failed.
Product code
KWA Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)
Reason for recall
The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Biomet Orthopedics sent an URGENT MEDICAL DEVICE RECALL NOTICE DATED April 14, 2011, to all affected customers. The notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: Immediately locate and remove the affected product. Follow the instructions on the enclosed "FAX Back Response Form" and fax a copy of the Response Form to 574-372-1683 prior to return of product. Customers were to confirm receipt of the notice by calling 800-348-9500, ext 3755 or 3756. For any questions regarding this recall call 574-371-3755 or 574-371-3756.
Quantity in commerce
146
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Australia, Netherlands and Japan.
Product codes and lots
Lot numbers M048180, M061180, M061190, M276160, M276170, M276180, M299730, M383260, M383290, M383300, M874150, M903110, M903160, M903180, M903190, M903200, M903220, M903230, and M048200.