Z-2415-2011

Pinnacle Destination Guiding Sheath, 6 French, 90 cm, Straight Tip, Tuohy-Borst Valve, Lot MD24, MD31, ME07...

Terumo Medical Corp. / initiated 2010-06-24 / Terminated / root cause: manufacturing process

Terumo Medical Corporation began a recall of Pinnacle Destination Guiding Sheath, 6 French, 90 cm, Straight Tip, Tuohy-Borst Valve, Lot MD24, MD31, ME07, & ME14. 510(k) K012812 Guiding Sheath is used to... on . The reason FDA records is "Molding defect in destination product that render the device unusable". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2415-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Medical Corp.
Product
Pinnacle Destination Guiding Sheath, 6 French, 90 cm, Straight Tip, Tuohy-Borst Valve, Lot MD24, MD31, ME07, & ME14. 510(k) K012812 Guiding Sheath is used to facilitate the introduction of interventional and diagnostic devices into the human vasculature, including but not limited to the renal arteries for the Renal Guiding Sheath and the carotid arteries for the Carotid Guiding Sheath.
Product code
DYB Introducer, Catheter
Reason for recall
Molding defect in destination product that render the device unusable.
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
Terumo Medical Corporation sent an Urgent Product Bulletin Recall letter dated June 24, 2010, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Terumo requested customers to immediately discontinue use, and quarantine any available stock of the listed product codes. In order to facilitate the exchange of these devices, a Terumo Territory Manager contacted all customers by July 9th to confirm receipt of the Terumo Destination Bulletin and scheduled an appointment to visit each customer in person by July 31st to personally manage the exchange. For questions or for further information, call (800) 283-7866 and ask to speak with the Product Recall Coordinator.
Quantity in commerce
214 units
Distribution
Worldwide Distribution - USA including AL, AK, AZ, TX, LA, MS, MN, MO, CA, GA, FL, HI, KS, IA, ID, IL, TN, NM, OR, OK, WA, and WI and the countries of Belgium, Brazil, and Panama