Z-2710-2011

The following kits contain the UROLOGIX, Coolant Bag, CA1111B, Sterile EO, Manufactured for: UROLOGIX, Inc...

Urologix, LLC / initiated 2011-04-28 / Terminated / root cause: manufacturing process

Urologix, Inc. began a recall of The following kits contain the UROLOGIX, Coolant Bag, CA1111B, Sterile EO, Manufactured for: UROLOGIX, Inc., 14405 Twenty-first Avenue North, Minneapolis, MN... on . The reason FDA records is "Urologix is recalling certain lots of Targis System Coolant Bag devices distributed over a two week period. The devices are packaged into disposable Procedure Kits containing a microwave catheter, a Coolant Bag and a...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2710-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Urologix, LLC
Product
The following kits contain the UROLOGIX, Coolant Bag, CA1111B, Sterile EO, Manufactured for: UROLOGIX, Inc., 14405 Twenty-first Avenue North, Minneapolis, MN 55447. UROLOGIX, CTC Advance Short, (2.5-3.5 cm), PT1034, Sterile EO,Manufactured by: UROLOGIX, Inc., 14405 Twenty-first Avenue North, Minneapolis, MN 55447. CTC Advance Standard , (3.0-5.0 cm), PT1036, Sterile EO,, Manufactured by: UROLOGIX, Inc., 14405 Twenty-first Avenue North, Minneapolis, MN 55447. CTC Advance Long, (4.5+ cm), PT1038, Sterile EO,Manufactured by: UROLOGIX, Inc., 14405 Twenty-first Avenue North, Minneapolis, MN 55447. Targis Standard, (3.0-5.0 cm), PT1013, Sterile EO,Manufactured by: UROLOGIX, Inc., 14405 Twenty-first Avenue North, Minneapolis, MN 55447. The Urologix Targis system is a non-surgical device intended to relieve symptoms and obstruction associated with Benign Prostatic Hyperplasia (BPH) and is indicated for men with prostatic urethral lengths greater than 2.5 cm.
Product code
MEQ System, Hyperthermia, Rf/Microwave (Benign Prostatic Hyperplasia),Thermotherapy
Reason for recall
Urologix is recalling certain lots of Targis System Coolant Bag devices distributed over a two week period. The devices are packaged into disposable Procedure Kits containing a microwave catheter, a Coolant Bag and a RTU Plus balloon. The reason for this Recall involves three lots of Coolant Bags that contain an increased risk of leakage of water at the Sensor Module. This leak is detectabl
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Urologix, Inc. sent an "IMPORTANT RECALL INFORMATION" letter dated April 28, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs the customers to identify the affected kit lots, remove the Coolant Bag pouch(s) and to complete and return a Return Material Authorization form with the product(s). Additionally, an enclosed Customer Acknowledgement Form should be completed and faxed to 763-475-1443. Customers with questions should contact the Director of Regulatory and Quality at 763-475-1400.
Quantity in commerce
119
Distribution
Nationwide Distribution--including the states of AL, CA, CO, FL, GA, MA, MI, MN, NJ, NY, OR, PA, TN, WV, and including Puerto Rico.
Product codes and lots
Catalog # / Lot No.  PT1034 / AHP0333111A, AHP040411A, AHP040411B, AHP040611A, AHP040711A, AHP040811A, AHP041211A  PT1036 / ACP033011B, ACP040111A, ACP040111B, ACP040611A, ACP040711A, ACP040711B, ACP040811A, ACP041211A  PT1038 / ALP033111A, ALP040411A  PT1013 / KP040711A  CA1111B / C900284, C900285 and C900286  C900284, C900285 and C900286  C900284 and C900285  C900285