Ambu Inc.
Columbia, MD, US
AMBU INC. appears in FDA device records in Columbia, MD. FDA records list no 510(k) clearances, no PMA records and one registered establishment. As of , FDA records show 11 recalls for this company, and at least one establishment registration is current.
FDA records held for this company
FDA records under these names
1 nameFDA records resolved to Ambu Inc. carry these names as well, as FDA wrote them. Each is the firm named on a registration, a 510(k), a PMA, a recall or an enforcement report that was joined to this company on a shared FDA identifier or on a matching name.
- Ambu Incorporatedrecall
510(k) clearance history
No 510(k) clearance records for this company in the ingested FDA data.
No 510(k) clearance records in the ingested FDA data.
Registered establishments
| Registration | FEI | Establishment | Address | Registration expires | Product codes listed |
|---|---|---|---|---|---|
| 1220828 | 1121788 | AMBU INC. | 6721 Columbia Gateway Drive, Suite 200, Columbia, MD 21046 US | 2026 | No device listings on this registration |
Registration numbers link to the FDA registration and listing record for that establishment.
FDA product codes
No FDA product codes recorded against this company.
Recall records
| Recall | Product | Reason | Classification | Initiated | Status |
|---|---|---|---|---|---|
| Z-2181-2025 | Ambu SPUR II Labeled as the following: 1 SPUR II Adult Resuscitator, Adult, Catalog Number: 523211000 2 SPUR II Adult Resuscitator w/Mercury Filter, Adult, Catalog Number: 523611051 3 SPUR II Adult Resuscitator, Single Patient Use Resuscitator, Adult, Catalog Number: 523611057 4 SPUR II Adult Resuscitator, Single Patient Use Resuscitator, Adult, Catalog Number: 524611000 5 SPUR II Adult Resuscitator, Pop-off open 40 ,w/PEEP Valve 20, Adult, Catalog Number: 524611001 6 SPUR II Adult Resuscitator w/PEEP Valve 20, Adult, Catalog Number: 524611011 7 SPUR II Adult Resuscitator w/Pressure Limiting Valve, Adult, Catalog Number: 524611031 8 SPUR II Adult Resuscitator w/Pressure Limiting Valve, Adult, Catalog Number: 524611047 9 SPUR II Adult Resuscitator, Single Patient Use Resuscitator, Adult, Catalog Number: 524611051 10 SPUR II Pediatric Resuscitator w/10 O2 tube & Manometer, Catalog Number: 530200016 11 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530212000 12 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530212001 13 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530213000 14 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530213001 15 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530213011 16 SPUR II Pediatric Resuscitator w/Manometer, Pediatric, Catalog Number: 530213030 17 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530213031 18 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530213048 19 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530214000 20 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530613000 21 SPUR II Pediatric Resuscitator w/Manometer & PEEP Valve 20, Pediatric, Catalog Number: 530613031 22 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530613071 23 SPUR II Pediatric Resuscitator w/PEEP Valve 20,Manometer & CO2 Detector, Pediatric, Catalog Number: 530613831 24 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530614017 25 SPUR II Pediatric Resuscitator w/Manometer, Pediatric, Catalog Number: 530614030 26 SPUR II Pediatric Resuscitator w/CO2 Detector, Pediatric, Catalog Number: 530614800 27 SPUR II Pediatric Resuscitator, INF& TOD W/Manometer, Pediatric, Catalog Number: 530615030 28 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530619031 29 SPUR II Pediatric Resuscitator w/Expiratory Filter, Manometer PEEP Valve, Pediatric, Catalog Number: 531600051 30 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 531613000 31 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 531613001 32 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 531613010 33 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 531613031 34 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 531613047 35 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number:531613051 36 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 531614026 37 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 531638000 38 SPUR II Adult Resuscitator, Single Patient Use Resuscitator, Adult, Catalog Number: 523611051E 39 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number:530213000B The Ambu SPUR II Resuscitator is a single patient use resuscitator intended for pulmonary resus | Potential for the manometer port being blocked rendering the manometer non-functional. | Class I | Open, Classified | |
| Z-1723-2025 | Ambu aScope" 5 Broncho HD 5.6/2.8 Sampler Set. Model Number: 622002000US | Incorrect labeling in which the front red pouch label did not match the actual size of the medical device. The label on the front of the scope incorrectly states 5.0/2.2 while the back label, shipping box and product inside correctly states 5.6/2.8. | Class II | Open, Classified | |
| Z-0261-2025 | Ambu VivaSight 2 DLT Endobronchial tube, 35 FR (Catalog Number 412351000), 37 Fr (Catalog Number 412371000), 39 Fr (Catalog Number 412391000), 41 Fr (Catalog Number 412411000) | Ambu has received complaints on Ambu VivaSight 2 DLTs regarding hyper angulation of the distal end. The hyper angulation can lead to an increased risk of complications during intubation and potential airway injury. | Class II | Open, Classified | |
| Z-0714-2024 | Ambu aView 2 A Monitor, Catalog numbers 405011000 and 405011000US2 | Instructions for Use have been updated to add the following warning: Use only M4 screws with the length of 14 16 mm when mounting Ambu aView 2 Advance on a VESA interface. Using longer screw lengths will penetrate the lithium-ion battery and result in a fire hazard and battery leakage which can cause severe burns, smoke inhalation and skin irritation. Using shorter screw lengths could result in unsecure device fastening. | Class II | Open, Classified | |
| Z-2628-2023 | Ambu aView 2 A Monitor-The Ambu aView 2 Advance is intended to display live imaging data from compatible Ambu visualization devices, portable endoscopy system. Catalog No: 405011000 | Multiple impacts, for example from being dropped, can cause to the Ambu aView 2 Advance to short-circuit and under special circumstances make the lithium-ion battery in the device combust leading to smoke and flames | Class II | Open, Classified | |
| Z-1420-2022 | Ambu VivaSight 2 DLT; endobronchial tube system used for verifying tube placement and repositioning. Catalog numbers 412351000, 412371000, 412391000, 412411000 | Firm has received complaints on the device concerning leak/rupture of either the bronchial or tracheal cuff. Leakage during procedure could lead to loss of secure airway and will require re-intubation of patient. | Class II | Open, Classified | |
| Z-1236-2016 | Ambu, Aura Gain - Single Use Laryngeal Mask- Sterile for use as an alternative to a face mask for achieving and maintaining control of the airway during routine emergency anesthetic procedures. | Ambu, Aura Gain laryngeal Mask, was found to have lack of Sterility Assurance. The firm discovered the problem during final quality control at manufacturing site. Damage to the sterile barrier (the pouch) may compromise a medical device's sterility,leading to contamination of the product and,ultimately,patient complications. | Class II | Terminated | |
| Z-0536-2008 | Ambu, Positive End Expiratory Pressure (PEEP) Valves, PEEP Valve Spare Part, REF 137 000 501, Manufacturer: Ambu International A/S, DK2750 Ballerup USA: Ambu Inc. (Ambu USA) Linthicum, MD 21090-1356 USA, Made in Denmark | Lower pressure than expected: Medical device for respiratory care may leak and not register accurate flow settings that affect patient respiration. | Terminated | ||
| Z-0535-2008 | Ambu, Positive End Expiratory Pressure (PEEP) Valves, PEEP Valve 10, 22/18 mm, 30mm connector, REF 137 001 000, Manufacturer: Ambu International A/S, DK2750 Ballerup USA: Ambu Inc. (Ambu USA) Linthicum, MD 21090-1356 USA, Made in Denmark | Lower pressure than expected: Medical device for respiratory care may leak and not register accurate flow settings that affect patient respiration. | Terminated | ||
| Z-0534-2008 | Ambu, Positive End Expiratory Pressure (PEEP) Valves, PEEP Valve 10, 22/18 mm, REF 000 138 000, Manufacturer: Ambu International A/S, DK2750 Ballerup USA: Ambu Inc. (Ambu USA) Linthicum, MD 21090-1356 USA, Made in Denmark | Lower pressure than expected: Medical device for respiratory care may leak and not register accurate flow settings that affect patient respiration. | Terminated | ||
| Z-0533-2008 | Ambu, Positive End Expiratory Pressure (PEEP) Valves, PEEP Valve 10, 30 mm, REF 000 137 000, Manufacturer: Ambu International A/S, DK2750 Ballerup USA: Ambu Inc. (Ambu USA) Linthicum, MD 21090-1356 USA, Made in Denmark | Lower pressure than expected: Medical device for respiratory care may leak and not register accurate flow settings that affect patient respiration. | Terminated |
Verification record
There is no verification record for this company. Most companies on this site have none: a record exists only where a company asked to be verified and MDM ran the checks, so the absence of one says nothing about the company.
Where there is a record, this is where it appears: what was checked, when it was checked, and the current status. An MDM Verified badge on another website links to a verify address on this domain that resolves to this section. A badge that does not resolve here is not ours.
MDM coverage
MDM coverage of this company lists here, newest first. Submit news.
Is this your company?
This page is built from public FDA records. Company pages are free and factual, and stay free whether or not anyone claims them. Membership never changes editorial coverage or vetting outcomes.
Claiming lets Ambu Inc. edit what it says about itself and list the categories it works in. It grants no badge: claiming is not verification.
Data as of , FDA export datePage updated Source: openFDA, public domain