Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0534-2008

Ambu, Positive End Expiratory Pressure (PEEP) Valves, PEEP Valve 10, 22/18 mm, REF 000 138 000, Manufacturer...

Ambu Inc. / initiated 2007-08-21 / Terminated / root cause: manufacturing process

Ambu Incorporated began a recall of Ambu, Positive End Expiratory Pressure (PEEP) Valves, PEEP Valve 10, 22/18 mm, REF 000 138 000, Manufacturer: Ambu International A/S, DK2750 Ballerup USA: Ambu... on . The reason FDA records is "Lower pressure than expected: Medical device for respiratory care may leak and not register accurate flow settings that affect patient respiration". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0534-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ambu Inc.
Product
Ambu, Positive End Expiratory Pressure (PEEP) Valves, PEEP Valve 10, 22/18 mm, REF 000 138 000, Manufacturer: Ambu International A/S, DK2750 Ballerup USA: Ambu Inc. (Ambu USA) Linthicum, MD 21090-1356 USA, Made in Denmark
Product code
BYE Attachment, Breathing, Positive End Expiratory Pressure
Reason for recall
Lower pressure than expected: Medical device for respiratory care may leak and not register accurate flow settings that affect patient respiration.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Ambu notified consignees with a letter dated 08/21/07 and flagged as Trouble Shooting Guide. The letter advised that some valves do not hold the set PEEP level under certain conditions and consignees were further advised to consider the guideline as an addendum to existing directions for use. Consignees were instructed to return (to Ambu) for replacement the identified lots listed in the notification and confirm receipt of the advisory by e-m or fax. Ambu subsequently identified additional lots of valves and replacement valve parts for recall and notified all consignees by "Urgent Device Recall" letter dated 11/29/07. The letter advised healthcare practitioners of the impact on patient treatment that may result in less than effective treatment. Consignees were further advised to cease use of the device and return for replacement.
Quantity in commerce
197 total
Distribution
Nationwide.
Product codes and lots
Catalog Number: A000138000, Lot Number:  010017.