Biocare Medical LLC

Pacheco, CA, US

BIOCARE MEDICAL, LLC appears in FDA device records in Pacheco, CA. FDA records list no 510(k) clearances, no PMA records and one registered establishment. As of , FDA records show 11 recalls for this company, and at least one establishment registration is current.

FDA records held for this company

510(k) clearances
0
PMA records
0
Recall records
11
Registered establishments
1

510(k) clearance history

No 510(k) clearance records for this company in the ingested FDA data.

No 510(k) clearance records in the ingested FDA data.

Registered establishments

FDA registered establishments for Biocare Medical LLC, with the product codes listed at each
RegistrationFEIEstablishmentAddressRegistration expiresProduct codes listed
30041403933004140393BIOCARE MEDICAL, LLC60 Berry Drive, Pacheco, CA 94553 US2026and 5 more, listed in full below

Registration numbers link to the FDA registration and listing record for that establishment.

FDA product codes

Recall records

Device recall records naming this company, most recent first
RecallProductReasonClassificationInitiatedStatus
Z-0315-2025intelliPATH Universal HRP Detection Kit, REF: IPK5011G80, containing DAB Buffer, Catalog Number: IPBF5009G20; ONCORE Pro DAB Chromogen, REF: OPRI6056KT180, containing DAB Buffer, Catalog Number: OPRI6055T60; IP DAB Chromogen Kit 80 ml, REF: IPK5010G80, containing DAB Buffer, Catalog Number: IPBF5009G20; DAB kit, REF: OPRI6056KT180, containing DAB Buffer, Catalog Number: OPRI6055T60 Folate Receptor alpha IHC Assay Kit, REF: IPI4006KG10, containing DAB Buffer, Catalog Number: IPBF5009G20Kits containing the DAB chromogen buffer, intended for use in either manual or automated Immunohistochemistry (IHC) staining protocols, may produce weak to no staining, which may lead to test failure requiring the test to be repeated.Class IIOpen, Classified
Z-1333-2021BIOCARE Medical , IntelliPath FLX, Automated Staining Instrument, Ref: " Model Number:IPS0001US and IPS0001INTL containing software version 3.5.3.1Their is a potential that the automated staining instrument with software version 3.5.3.1 may move with random speeds in random direction across its range of motion. This could result in the instrument hitting end stops or running into other components on the working deck. This could also result in wash buffer being dispensed on random locations.Class IITerminated
Z-2291-20204plus Streptavidin HRP Label, Conjugated Streptavidin Horseradish Peroxidase, Detection Component, Catalog Number: HP604HFirm has identified product has no staining. If used, may result in invalid test results when used with controls in the clinical setting. Positive control would observe no staining. The clinical impact may result in the consumption of patient tissue sample and a delay in diagnosis.Class IITerminated
Z-2360-2018Vulcan Fast Red Chromogen Kit 2, part of an immunohistochemistry (IHC) procedure using an alkaline phosphatase (AP) detection system. Catalog numbers FR805-5L and FR805SChromogen failed to develop due to a stability issue caused by environmental factors with a raw material in the Vulcan Fast Red Buffer.Class IITerminated
Z-3198-2018URO-3 Triple Stain (CD44 + p53) with CK20, Catalog #: PM370TSAA, Package Type: Bag, which contains components: a) CD44 + p53, Catalog #: PM373DSAA, Package Type: Bottle, and b) CK20, Catalog #: PM062AA, Package Type: Bottle Product Usage: URO-3 Triple Stain (CD44 + p53) with CK20 is a cocktail of mouse monoclonal and rabbit monoclonal antibodies that is intended for laboratory use in the qualitative identification of CD44, p53 and CK20 proteins by immunohistochemistry (IHC) in formalin-fixed paraffin-embedded (FFPE) human tissues, primarily for bladder cancer characterization. The clinical interpretation of any staining or its absence should be complemented by morphological studies using proper controls and should be evaluated within the context of the patient s clinical history and other diagnostic tests by a qualified pathologist. The URO-3 Triple Stain is an adjunct immunohistochemistry test to assist a qualified pathologist in differentiating urothelial reactive atypia from CIS (carcinoma in situ) in suspected bladder lesions. The URO-3 staining pattern will be used as supporting data along with morphological and patient history context for the qualified pathologist to render a diagnosis.Immunohistochemistry device kits found to display incorrect staining pattern due to mislabeling of two components.Class IITerminated
Z-1523-2016Mach 4 Mouse Probe vial within the Mach 4 Universal HRP-Polymer Kit; Catalog number M4U534L; For In Vitro Use Immunology: MACH 4 Universal HRP-Polymer is a horseradish peroxidase (HRP)-antibody conjugate system intended for use in the detection of mouse lgG and lgM, and rabbit IgG primary antibodies on formalin-fixed, paraffin-embedded (FFPE) tissues in an immunohistochemistry (IHC) procedure. The clinical interpretation of any staining or its absence should be complemented by morphological studies and proper controls and should be evaluated within the context of the patient's clinical history and other diagnostic tests by a qualified pathologist.Biocare Medical has identified a single lot (082715) of its Mach 4 HRP Polymer kit in which some few Mach 4 Mouse Probe vials may have been mislabeled as HRP-Polymer.Class IIITerminated
Z-0252-2016Smooth Muscle Action (SMA) Immunohistochemistry reagent. Catalog numbers: CME305A-0.1ml; CME305B - 0.5 ml, PME305AA - 6.0 ml prediluted and ready to use. They are the same product with the same intended use. Chemistry - IVD Smooth Muscle Actin (SMA) antibody recognizes the alpha-smooth muscle isoform of actin. This antibody recognizes the alpha-smooth muscle isoform of actin. This MAb is reportedly useful for identifying tumors arising from smooth muscle and myoepithelial cells.Certain catalog numbers and lots of Smooth Muscle Actin may stain some cell types that are typically expected to be negative for SMA.Class IITerminated
Z-1676-2015D2-40 (Lymphatic Marker); For In Vitro Diagnostic; Catalog number: PM266AA Product Usage: In Vitro Diagnostic Use.: a mouse monoclonal antibody that is intended for laboratory use in the qualitative identification of O-linked sialoglyprotein protein by immunohistochemistry (ICH) in formalin fixed paraffin embedded human tissues.Firm discovered that incorrect Instructions for Use were shipped with one lot of D2-40.Class IIITerminated
Z-1535-2015CD99 Catalog number PM008 AA; For In Vitro Diagnostic Use. Product Usage: Provides adjunctive diagnostic information following the primary diagnosis of a neoplasm by conventional histopathology methods.A drop in staining intensity over time has been observed. These lots may not meet their labeled shelf life.Class IITerminated
Z-0965-2015MUM-1 (Multiple myeloma oncogene-1), catalog number PRM352 AA; Product Usage: For In Vitro Diagnostic Use. This antibody may be used as a tool for the identification and the sub classification of lymphoid malignancies : Multiple myeloma oncogene-1 (MUM-1) is a protein encoded by MUM-1 gene. MUM-1 protein is expressed in the nuclei and cytoplasm of plasma cells and a small percentage of germinal center (GC) B-cells located in the light zone. MUM-1 [BC5] labels MUM-1protein in centrocytes and their progeny, plasma cells, activated T-cells and a wide spectrum of hematolymphoid neoplasms derived from these cells. MUM-1 has been reported to play an important role in mediating B-cell activation and differentiation.Three lots of MUM-1 may not meet their labeled shelf life, as a drop in staining intensity over time has been observed.Class IITerminated
Z-1392-2012BioCare Medical Vina Green Chromogen Kit; 25 mL, 100ml IVD. Kits include: Vina Green Chromogen 1. 0 mL; Vina Green Buffer, 25 mL and Vina Green Dropper Bottle. BioCare Medical, Concord, CA 94520. Vina Green is intended for both IHC and ISH applications including HPV, CMV, EBV, Kappa, TTF-1, Ki-67 and other targets/antigens such as blood and lymphatic vessels and basal and myoepithelial cells.Reports of sporadic failure of specific lots of Vina Green Chromogen Kit, either through crystal formation that obscures positive staining, color not right, or a complete fading of staining by the chromogen.Terminated

Verification record

There is no verification record for this company. Most companies on this site have none: a record exists only where a company asked to be verified and MDM ran the checks, so the absence of one says nothing about the company.

Where there is a record, this is where it appears: what was checked, when it was checked, and the current status. An MDM Verified badge on another website links to a verify address on this domain that resolves to this section. A badge that does not resolve here is not ours.

What MDM checks, and what it does not

MDM coverage

MDM coverage of this company lists here, newest first. Submit news.

Is this your company?

This page is built from public FDA records. Company pages are free and factual, and stay free whether or not anyone claims them. Membership never changes editorial coverage or vetting outcomes.

Claiming lets Biocare Medical LLC edit what it says about itself and list the categories it works in. It grants no badge: claiming is not verification.

Claim this page

Data as of , FDA export datePage updated Source: openFDA, public domain