Z-0315-2025Class II

intelliPATH Universal HRP Detection Kit, REF: IPK5011G80, containing DAB Buffer, Catalog Number: IPBF5009G20...

Biocare Medical LLC / initiated 2024-09-25 / Open, Classified / root cause: manufacturing process

Biocare Medical, LLC began a recall of intelliPATH Universal HRP Detection Kit, REF: IPK5011G80, containing DAB Buffer, Catalog Number: IPBF5009G20; ONCORE Pro DAB Chromogen, REF: OPRI6056KT180... on . FDA classified it as Class II and gives the reason as "Kits containing the DAB chromogen buffer, intended for use in either manual or automated Immunohistochemistry (IHC) staining protocols, may produce weak to no staining, which may lead to test failure requiring the test...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0315-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Biocare Medical LLC
Product
intelliPATH Universal HRP Detection Kit, REF: IPK5011G80, containing DAB Buffer, Catalog Number: IPBF5009G20; ONCORE Pro DAB Chromogen, REF: OPRI6056KT180, containing DAB Buffer, Catalog Number: OPRI6055T60; IP DAB Chromogen Kit 80 ml, REF: IPK5010G80, containing DAB Buffer, Catalog Number: IPBF5009G20; DAB kit, REF: OPRI6056KT180, containing DAB Buffer, Catalog Number: OPRI6055T60 Folate Receptor alpha IHC Assay Kit, REF: IPI4006KG10, containing DAB Buffer, Catalog Number: IPBF5009G20
Product code
PPM General Purpose Reagent
Reason for recall
Kits containing the DAB chromogen buffer, intended for use in either manual or automated Immunohistochemistry (IHC) staining protocols, may produce weak to no staining, which may lead to test failure requiring the test to be repeated.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On 9/25/24, recall notices were sent to customers and distributors who were asked to do the following: 1) Locate any affected kits in your possession and discontinue use. 2) Share this notification with all users within your facility and network to ensure all personnel who may use the kits are aware of this recall. 3) Discard any product remaining in your possession. 4) Complete and return the Response Form via email to [email protected] Distributors were asked to send the customer notice to their customers. Questions regarding this recall notification may be forwarded to [email protected] On 12/12/2024, recall notices were sent to customers containing additional affected lots.
Quantity in commerce
976
Distribution
US Nationwide distribution in the states of MA, AZ, CT, NY, FL, NC, IN, IL, MO, CO, TX, CA, MI, PR, PA, MD, AL, OH, KY.
Product codes and lots
REF/UDI-DI/Kit Lot(DAB Lot): IPK5011G80/00847627009005/042324(012424A),060523(051823A), 082923(062923A), 011624(072723A), 042324(012424A), 080524(032624A); OPRI6056KT180/00847627024985/020624(111623A),032224(111623A-3); IPK5010G80/00847627008992/071923(051823A), 091923(081023A), 112823(101923A), 100424(011524A), 041624(022624A); OPRI6056KT180/00847627024985/071924(061824A); IPI4006KG10/00847627008985/080323(062923A), 102723(081023A)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0315-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0315-2025Class IIOngoingVoluntary: Firm initiated