C P Medical Inc.
Portland, OR, US
C P Medical Inc. appears in FDA device records in Portland, OR. FDA records list no 510(k) clearances, no PMA records and no registered establishments. As of , FDA records show 9 recalls for this company.
FDA records held for this company
- 510(k) clearances
- 0
- PMA records
- 0
- Recall records
- 9
- Registered establishments
- 0
510(k) clearance history
No 510(k) clearance records for this company in the ingested FDA data.
No 510(k) clearance records in the ingested FDA data.
Registered establishments
No FDA establishment registrations recorded against this company.
FDA product codes
No FDA product codes recorded against this company.
Recall records
| Recall | Product | Reason | Classification | Initiated | Status |
|---|---|---|---|---|---|
| Z-0140-2015 | Monoswift PGCL [poly(glycolide-co-caprolactone)] Absorbable Surgical Suture; USP size 3/0. Product is labeled sterile. The product is sealed inside an inner foil pouch with an outside Tyvek pouch. It is an undyed, monofilament synthetic absorbable suture with needle size 3/0. | CP Medical Inc. is recalling five lots of Monoswift PGCL [poly(glycolide-co-caprolactone)] Absorbable Surgical Suture for failing to meet the minimum tensile strength as dictated by USP was not met over the length of the product's shelf life. | Class II | Terminated | |
| Z-1708-2014 | Monoswift PGCL [poly (glycolide-co-caprolactone)] Absorbable Surgical Suture; USP size 3/0 PGCL; part number L936. Product is labeled sterile. The product is sealed inside an inner foil pouch with an outside Tyvek pouch. It is an undyed, monofilament synthetic absorbable suture with needle, size 3/0 (2.0 metric), PS-1 Needle | CP Medical, Inc., is recalling one lot of Monoswift PGCL [poly (glycolide-co-caprolactone)] Absorbable Sutures for failing to meet the USP minimum average value for knot pull (tensile) strength. | Class II | Terminated | |
| Z-3177-2011 | Monomid Non-Absorbable Suture are blue, monofilament synthetic absorbable suture, with different sizes and different needle sizes. The products are labeled in parts. These sutures are intended for use in general soft tissue approximation and/or ligation including use in cardiovascular, ophthalmic, and neurological surgery. | Unapproved dye was used in manufacturing the non-absorbable blue nylon suture material. | Terminated | ||
| Z-0260-2012 | Product is a synthetic absorbable suture with needle and labeled as sterile. The product is sealed inside an inner foil pouch with an outside Tyvek pouch. The outer sterile barrier seal is intact. The product is boxed and shrink wrapped for protection against contamination. The product is a Violet, monofilament synthetic absorbable suture with needle, size 3/0 (2.0 Metric), NFS-2 Needle (Product code 398A-Polyglycolic Acid) or size 2/0 (3.0 Metric) NFS-1 Needle. (Product code 453A-Polyglycolic Acid). Product is labeled in part: "***NFS-1***CUTTING***Coated POLYGLYCOLIC ACID Violet (PGA) Braided Absorbable Suture STERILE***Manufacturer CP Medical***". Intended for use in general soft tissue approximation and/or ligation including use in pediatric cardiovascular tissue where growth is expected to occur, and ophthalmic surgery. The suture is not recommended for adult cardiac tissue, microsurgery or neural tissue. The absorbable sutures are useful where extended wound support is desirable. | The seal integrity of the outer product pouch may be compromised due to partial delamination of the two layers that make up part of the pouch. | Terminated | ||
| Z-0259-2012 | Product is a synthetic absorbable suture with needle and labeled as sterile. The product is sealed inside an inner foil pouch with an outside Tyvek pouch. The outer sterile barrier seal is intact. The product is boxed and shrink wrapped for protection against contamination. The product is a Violet, monofilament synthetic absorbable suture with needle, size 4/0 (1.5 Metric), NFS-2 Needle (Product code M397-Polydioxanone) or size 3/0 (2.0 Metric) (Product code M398 {Butler code 029254}). Product is labeled in part: "***POLYDIOANONE Violet (PDO) Monofilament Absorbable Suture***". Intended for use in general soft tissue approximation and/or ligation including use in pediatric cardiovascular tissue where growth is expected to occur, and ophthalmic surgery. The suture is not recommended for adult cardiac tissue, microsurgery or neural tissue. The absorbable sutures are useful where extended wound support is desirable. | The seal integrity of the outer product pouch may be compromised due to partial delamination of the two layers that make up part of the pouch. | Terminated | ||
| Z-2582-2010 | Product is labeled in part: "CP-SM397A-02...Mono-Dox Violet Monofilament Polydioxanone Synthetic Absorbable Sterile Suture. Do Not Resterilize.... 4/0 (1.5 metric) 30" (75cm) Rx Only...LOT L0535440 EXP 2009-11...4/0 M397A...Mono-Dox CUTTING NFS-2 NEEDLE...Manufactured by: CP Medical, Portland, OR USA". Intended for use in general soft tissue approximation and/or ligation including use in pediatric cardiovascular tissue where growth is expected to occur and ophthalmic surgery. | Seal integrity of inner foil pouch of the Mono-Dox Synthetic Absorbable Suture may be compromised. | Terminated | ||
| Z-0809-2007 | Product is an EO Sterile, Polydioxanone Violet monofilament synthetic absorbable surgical suture with needle, size 3/0 (2.0 Metric), NFS-2, 19 mm Needle. Product is sealed inside an inner foil pouch with an outside tyvek pouch. The outer tyvek pouch has the defective seal. No suture is exposed or visible through the foil wrap which must be torn to have access to the suture and needle. Product is boxed (12 to a box) and shrink wrapped. Product is labeled STERILE. Box Label includes: '3/0 (2.0 metric) MONO-DOX Violet Monofilament Polydioxanone Suture USP M398A CP-B-M398A-04 Rx Only NFS-2 19 mm 3/8 CUTTING Manufactured by: CP Medical Portland, OR USA 30' (75 cm) One Dozen Sterile Synthetic Absorbable Suture '. Packet Label includes: '3/0 (2.0 metric) 398 NFS-2 19mm 3/8 CUTTING 30' (75cm) POLYDIOXANONE Violet (PDO) Monofilament Synthetic Absorbable Suture STERILE EO Manufacturer: CP Medical Portland, OR USA' . Product is also distributed under three private labels, Webster Veterinary, Butler, and Phoenix Pharmaceutical, Inc. Webster Veterinary Box Label includes: '3/0 (2.0 metric) WEBMAX Violet Monofilament Polydioxanone Suture Distributed by: WEBSTER VETERINARY 07-836-4288(M398A) 26-BM398A-01 Vet Use Only CUTTING NFS-2 30' (75 cm) 3/8 19mm'. Webster Veterinary Packet Label includes: '3/0 (2.0 metric) 07-836-4288(M398A) WEBMAX Violet Monofilament Polydioxanone Absorbable Sterile Suture 30' (75cm) Vet Use Only Distributed by: Webster Veterinary Supply, Inc. CUTTING NFS-2 3/8 19mm 26-SM398A-01'. Butler Box label includes: 'Butler 3/0 (2.0 metric) MONOSORB Violet Monofilament Polydioxanone Suture 029254 Vet Use Only NFS-2 CUTTING 30' (75 cm) 3/8 19mm 01-BM398A-02 SUTUREVET'. Butler Packet Label includes: '3/0 (2.0 metric) NFS-2 19mm 3/8 CUTTING 30' (75cm) 029254 MONOSORB Violet Monofilament Prolydioxanone Synthetic Absorbable Suture STERILE EO Distributed Exclusively by: Butler Animal Health Supply'. Phoenix Pharmaceutical Box label includes: '3/0 (2.0 metric) 0398 Poly-Dox Sutures VIOLET MONOFILAMENT (Polydioxanone) Suture Sterile Synthetic Absorbable Suture 30' (75cm) Vet Use Only Cutting PFS-2 3/8 19mm Distributed By: PHOENIX PHARMACEUTICAL, INC.'. Phoenix Pharmaceutical Packet label includes: '3/0 (2.0 metric) 0398 Poly-Dox Sutures VIOLET MONOFILAMENT (Polydioxanone) Suture 30' (75cm) Sterile, Synthetic Absorbable Suture Veterinary Use Only Cutting PFS-2 3/8 19mm Distributed By: PHOENIX PHARMACEUTICAL, INC.'. | Potential for sterility of product to be compromised. Suture package seals incomplete or missing. | Terminated | ||
| Z-0808-2007 | Product is an EO Sterile, Blue Polypropylene monofilament synthetic nonabsorbable surgical suture with needle, size 3/0 (2.0 Metric), NFS-2, 19 mm Needle. Product is wrapped in a five-fold card with inserted flap ends (overwrap). A label is placed on the overwrap and the packet is then inserted into a tyvek/mylar pouch. One end of the overwrap card is printed with 'Open'. This end is positioned in the pouch so it is closest to the peel-open end. No suture is exposed or visible through the overwrap which must be torn to have access to the suture and needle. Product is boxed (12 to a box) and shrink wrapped. Product is labeled STERILE. Box Label includes: '3/0 (2.0 metric) POLYPRO Blue Monofilament Polypropylene Nonabsorbable Suture 8665P Rx Only NFS-2 19 mm 3/8 CUTTING 18' (45 cm) One Dozen Sterile Manufactured by: CP Medical Portland, OR USA' . Side of box has an adhesive sticker label which includes: 'Vet Use Only'. Packet Label includes: '3/0 (2.0 metric) 8665 NFS-2 19mm 3/8 TAPER POLYPROPYLENE Blue Polypropylene Monofilament Synthetic Nonabsorbable Suture 18' (45 cm) STERILE EO Manufacturer: CP Medical Portland, OR USA' . Product is also distributed under two private labels, Webster Veterinary and Butler. Webster Veterinary Box Label includes: '3/0 (2.0 metric) WEBPRO Blue Monofilament Polypropylene Nonabsorbable Suture Distributed by: WEBSTER VETERINARY One Dozen Sterile 07-836-4643(8665P) 26-B8665P-02 Vet Use Only CUTTING NFS-2 18' (45 cm) 3/8 19mm'. Webster Veterinary Packet Label includes: '3/0 (2.0 metric) WEBPRO Blue Monofilament Polypropylene Nonabsorbable Sterile Suture Vet Use Only Distributed by: Webster Veterinary Supply, Inc. 07-836-4643(8665P) 26-B8665P-02 NFS-2 18' (45 cm) CUTTING 3/8 19mm'. Butler Box label includes: 'Butler 3/0 (2.0 metric) PROMEND Blue Monofilament Polypropelene 032002 Vet Use Only NFS-2 CUTTING 18' (45 cm) 3/8 19mm 01-B8665P-01 SUTUREVET'. Butler Packet Label includes: '3/0 (2.0 metric) NFS-2 19mm 3/8 CUTTING 18' (45cm) 032002 PROMEND Blue Propylene Monofilament Synthetic Nonabsorbable Suture STERILE EO Distributed Exclusively by: Butler Animal Health Supply'. | Potential for sterility of product to be compromised. Suture package seals incomplete or missing. | Terminated | ||
| Z-0807-2007 | Product is an EO Sterile, Polyester braided and coated nonabsorbable surgical suture with double needle, size 2/0 (3.0 Metric), NSH 26 mm TAPER Needle. Product is wrapped in a five-fold card with inserted flap ends (overwrap). A label is placed on the overwrap and the packet is then inserted into a tyvek/mylar pouch. One end of the overwrap card is printed with 'Open'. This end is positioned in the pouch so it is closest to the peel-open end. No suture is exposed or visible through the overwrap which must be torn to have access to the suture and needle. Product is boxed (12 to a box) and shrink wrapped. Product overwrap label includes: '2/0 (3.0 metric) CP523A Polybond Green Braided, Polyester Coated Nonabsorbable Sterile Suture Rx Only TAPER NSH 36' (90cm) 1/2 26mm Manufactured by CP Medical Portland, OR USA'. Product box label includes: '2/0 (3.0 metric) CP523A NSH 26 mm 1/2 TAPER POLYBOND Green Polyester Braided Coated Nonabsorbable 36' (90cm) STERILE EO Manufacturer CP Medical Portland, OR USA' . Product was distributed to single consignee, a distributor in South Africa. | Potential for sterility of product to be compromised. Suture package seals incomplete or missing. | Terminated |
Verification record
There is no verification record for this company. Most companies on this site have none: a record exists only where a company asked to be verified and MDM ran the checks, so the absence of one says nothing about the company.
Where there is a record, this is where it appears: what was checked, when it was checked, and the current status. An MDM Verified badge on another website links to a verify address on this domain that resolves to this section. A badge that does not resolve here is not ours.
MDM coverage
MDM coverage of this company lists here, newest first. Submit news.
Is this your company?
This page is built from public FDA records. Company pages are free and factual, and stay free whether or not anyone claims them. Membership never changes editorial coverage or vetting outcomes.
Claiming lets C P Medical Inc. edit what it says about itself and list the categories it works in. It grants no badge: claiming is not verification.
Data as of , FDA export datePage updated Source: openFDA, public domain