Carl Zeiss Microscopy, LLC
White Plains, NY, US
CARL ZEISS MICROSCOPY, LLC appears in FDA device records in White Plains, NY. FDA records list no 510(k) clearances, no PMA records and 2 registered establishments. As of , FDA records show 8 recalls for this company, and at least one establishment registration is current.
FDA records held for this company
FDA records under these names
1 nameFDA records resolved to Carl Zeiss Microscopy, LLC carry these names as well, as FDA wrote them. Each is the firm named on a registration, a 510(k), a PMA, a recall or an enforcement report that was joined to this company on a shared FDA identifier or on a matching name.
- Zeiss, Carl Incenforcement report, recall
510(k) clearance history
No 510(k) clearance records for this company in the ingested FDA data.
No 510(k) clearance records in the ingested FDA data.
Registered establishments
| Registration | FEI | Establishment | Address | Registration expires | Product codes listed |
|---|---|---|---|---|---|
| 2431026 | 2431026 | CARL ZEISS MICROSCOPY, LLC | One North Broadway, White Plains, NY 10601 US | 2026 | No device listings on this registration |
| 8010666 | 3012769735 | Carl Zeiss Microscopy, GmbH | CARL ZEISS PROMENADE 10, Jena Thuringia 07745 DE | 2026 |
Registration numbers link to the FDA registration and listing record for that establishment.
FDA product codes
Recall records
| Recall | Product | Reason | Classification | Initiated | Status |
|---|---|---|---|---|---|
| Z-1465-2022 | ZEISS Axio Observer 5 microscope, Product material no. 431006-9901-000 | An incorrect Unique Device Identification (UDI) label was installed on some ZEISS Axio Observer 5 microscopes. | Class III | Terminated | |
| Z-1629-2020 | Component for Laser Scanning Microscopes (LSM 700, LSM 800 and LSM 900) | When the user tilts the transmitted light arm of the microscope to the end of its axis range, laser safety can be bypassed and result in emission of class lIlb laser radiation. This issue arises when a ZEISS inverted stage stand (Axio Observer or Axio Vert) is used in as a component of the LSM 700, 800 and 900 system | Class II | Terminated | |
| Z-2441-2018 | Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope | Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation. | Class II | Terminated | |
| Z-2058-2018 | Cell Observer SD, DirectFRAP, Laser TIRF 1, Laser TIRF 2, Laser TIRF3, LightSheet, Elyra, and multiple Laser Scanning Microscopes. Laser scanning microscopes are used in cell biology research. | In certain eyepiece configurations the laser shutter may not close completely, permitting reflected beams that may be greater than the Class I limit. | Class II | Terminated | |
| Z-0799-2014 | Compensator Slider Lambda 6x20 component (part number 473704-0000-000) utilized on all upright ZEISS microscopes (Axio Imager, Axio Scope. A1, Axio Lab. A1). Component used to determine composition of urinary calculi, urinary sediments, bio-crystallates and diagnosis of pathological characteristics in medical applications (e.g.: testing for gout or pseudogout); also used for non-medical scientific research. | Carl Zeiss Microscopy, LLC has recalled the Compensator Slider Lamda 6x20 component utilized on Standard Pathology Microscopes due to misalignment of the lambda plate. | Class II | Terminated | |
| Z-1744-2013 | Polarizer D Fixed w/Rotating Red 1 lambda plate. Utilized in upright Zeiss microscopes to aid in diagnosis by means of color variation. | Carl Zeiss Microscopy, LLC is recalling the Polarizer D component used in All upright ZEISS microscopes (Axio Imager, Axio Scope.A1, Axio Lab.A1) due to Misaligned Full-Wave (Lambda) Plate. | Class II | Terminated | |
| Z-1129-2012 | DirectFRAP (Direct Flourescence Recovery After Photobleaching) Imaging System is an extension unit for Axio Observer that is used for microscopic analyses of dynamic processes within living cells. The DirectFRAP module is a component used in conjunction with Laser Scanning Microscopes, Non-Medical Devices for Research Use. The DirectFRAP Imaging System is a Class 3B laser system. The DirectFRAP Imaging System is a laser manipulation slider which allows coupling of laser light simultaneously with the conventional fluorescence illumination into the reflected-light beam path of the microscope system. Depending on the application, the laser light is used for the targeted bleaching of fluorophores for FRAP experiments or for the conversion of fluorophores within the cell. | The manufacturer confirmed that the opto-semiconductor sensors can be triggered by exposure to incandescent lighting. The safety interlock is in place to prevent unintended exposure to Class 3B laser radiation. | Terminated | ||
| Z-1128-2012 | Carl Zeiss Microimaging Laser-TIRF (Total Internal Reflection Fluorescence) Imaging System is a component used in conjunction with Laser Scanning Microscopes, Non-Medical Devices for Research Use. The Laser-TIRF Imaging-System is a Class 3B laser system. The Laser TIRF Imaging System allows the visualization of substances or molecules being capable of fluorescing (excitable by light) by means of an electromagnetic (evanescent) field produced by total internal reflection. | The manufacturer confirmed that the opto-semiconductor sensors can be triggered by exposure to incandescent lighting. The safety interlock is in place to prevent unintended exposure to Class 3B laser radiation. | Terminated |
Verification record
There is no verification record for this company. Most companies on this site have none: a record exists only where a company asked to be verified and MDM ran the checks, so the absence of one says nothing about the company.
Where there is a record, this is where it appears: what was checked, when it was checked, and the current status. An MDM Verified badge on another website links to a verify address on this domain that resolves to this section. A badge that does not resolve here is not ours.
MDM coverage
MDM coverage of this company lists here, newest first. Submit news.
Is this your company?
This page is built from public FDA records. Company pages are free and factual, and stay free whether or not anyone claims them. Membership never changes editorial coverage or vetting outcomes.
Claiming lets Carl Zeiss Microscopy, LLC edit what it says about itself and list the categories it works in. It grants no badge: claiming is not verification.
Data as of , FDA export datePage updated Source: openFDA, public domain
