Polarizer D Fixed w/Rotating Red 1 lambda plate. Utilized in upright Zeiss microscopes to aid in diagnosis by...
Zeiss, Carl Inc began a recall of Polarizer D Fixed w/Rotating Red 1 lambda plate. Utilized in upright Zeiss microscopes to aid in diagnosis by means of color variation on . FDA classified it as Class II and gives the reason as "Carl Zeiss Microscopy, LLC is recalling the Polarizer D component used in All upright ZEISS microscopes (Axio Imager, Axio Scope.A1, Axio Lab.A1) due to Misaligned Full-Wave (Lambda) Plate". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1744-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Carl Zeiss Microscopy, LLC
- Product
- Polarizer D Fixed w/Rotating Red 1 lambda plate. Utilized in upright Zeiss microscopes to aid in diagnosis by means of color variation.
- Product code
- IBJ Light, Microscope
- Reason for recall
- Carl Zeiss Microscopy, LLC is recalling the Polarizer D component used in All upright ZEISS microscopes (Axio Imager, Axio Scope.A1, Axio Lab.A1) due to Misaligned Full-Wave (Lambda) Plate.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On May 28, 2013, Carl Zeiss Microscopy, LLC recalled a Polarizer Component due to Misaligned Full-Wave (Lambda) Plate. Carl Ziess Microscopy distributed notices to their customers via e-mail. Carl Zeiss Microscopy, LLC requests that all customers immediately discontinue the use of the component for any applications which rely on accurate color-change sequence results. Your service provider will contact you to schedule a visit in which a factory trained service technician will confirm the alignment of your Polarizer D component;s lambda plate. If it is defective, a new Polarizer D component will be installed. There will be no charge for this action. Should you have any questions related to this issue, please do not hesitate to the Carl Zeiss facility in New York by telephone 914-681-7815 or via email at [email protected]
- Quantity in commerce
- 10
- Distribution
- Distributed in the states of MI, TX, SD, FL, GA, NY, and NY.
- Product codes and lots
- Model # 445226-0000-000
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1744-2013 | Class II | Terminated | Voluntary: Firm initiated |