GE Medical Systems China Co., Ltd.

WUXI Jiangsu, CN

GE MEDICAL SYSTEMS (CHINA) CO., LTD. appears in FDA device records in WUXI Jiangsu, CN. FDA records list 13 510(k) clearances between and , no PMA records and one registered establishment. As of , FDA records show 27 recalls for this company, and at least one establishment registration is current.

FDA records held for this company

510(k) clearances
13
PMA records
0
Recall records
27
Registered establishments
1

510(k) clearance history

FDA records hold 13 510(k) clearances for GE Medical Systems China Co., Ltd. between 2009 and 2017. The busiest year on record is 2013, with 3. The most recent is 2017, with 1.

Clearances by year, as a table
510(k) clearance decisions for GE Medical Systems China Co., Ltd., by decision year
YearClearances
20092
20102
20111
20121
20133
20142
20151
20171
510(k) clearances on record, most recent first
K numberDeviceProduct codeDecisionDecision date
K171580Monitor B125, Monitor B105MHXSubstantially Equivalent
K151063Monitor B40MHXSubstantially Equivalent
K142288MAC 800 Resting ECG Analysis SystemDPSSubstantially Equivalent
K133576MONITOR B40MHXSubstantially Equivalent
K130584MONITOR B40MHXSubstantially Equivalent
K122253PROCARE MONITOR B20MHXSubstantially Equivalent
K123238ACHILLESMUASubstantially Equivalent
K120598PROCARE MONITOR B40MHXSubstantially Equivalent
K103633ACHILLESMUASubstantially Equivalent
K102256GE LOGIQ I/E & VIVID E COMPACT DIAGNOSTIC ULTRASOUNDIYNSubstantially Equivalent
K102113GE VENUE 40 COMPACT DIAGNOSTIC ULTRASOUNDIYNSubstantially Equivalent
K091374GE LOGIQ I/E & VIVID E COMPACT ULTRASOUNDIYOSubstantially Equivalent
K091164GE VENUE 40, MODEL 5324338IYNSubstantially Equivalent

Registered establishments

FDA registered establishments for GE Medical Systems China Co., Ltd., with the product codes listed at each
RegistrationFEIEstablishmentAddressRegistration expiresProduct codes listed
97106023003513129GE MEDICAL SYSTEMS (CHINA) CO., LTD.NO. 19 CHANGJIANG ROAD, WUXI Jiangsu 214028 CN2026and 47 more, listed in full below

Registration numbers link to the FDA registration and listing record for that establishment.

FDA product codes

Recall records

Showing 25 of 27
Device recall records naming this company, most recent first
RecallProductReasonClassificationInitiatedStatus
Z-2306-2026MAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. ElectrocardiographSoftware in development inadvertently made available for customer use. Use of this software version can result in clinical information being associated with the incorrect patient. This could impact clinical decision-making.Class IIOpen, Classified
Z-0915-2026GE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU), 2076139-001-SPotential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.Class IOpen, Classified
Z-0914-2026GE Healthcare Carestation anesthesia system, product number and REF numbers: Carestation 750 A1, REF 1012-9750-000; Carestation 750 A2, REF 1012-9750-002; Carestation 750c A1, REF 1012-9755-000; Products for which affected PMB can be used as a spare part: Carestation 750c A2, REF 1012-9755-002Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.Class IOpen, Classified
Z-0913-2026GE Healthcare Carestation anesthesia system, product number and REF numbers: Carestation 610 A1, REF 1012-9620-222; Carestation 620 A1 REF 1012-9620-000; Carestation 620 A1, REF 1012-9620-200; Carestation 620 A1, REF 1012-9620-202; Carestation 620 SE A1, REF 1012-9620-212; Carestation 630 A1, REF 1012-9650-222; Carestation 650 A1, REF 1012-9650-000; Carestation 650 A1, REF 1012-9650-200; Carestation 650 A1, REF 1012-9650-202; Carestation 650 SE A1, REF 1012-9650-212; Carestation 650c A1, REF 1012-9655-202; Anesthesia Gas Machine Products for which affected PMB can be used as a spare part: Carestation 620 A2, REF 1012-9620-002 00840682124546 Carestation 620 SE A2, REF 1012-9620-012 00195278569677 Carestation 650 A2, REF 1012-9650-002 00840682124560 Carestation 650 SE A2, REF 1012-9650-012 00195278569684 Carestation 650c A1, REF 1012-9655-000 00840682103954 Carestation 650c A2, REF 1012-9655-002 00840682124539 Carestation 650c A1, REF 1012-9655-200 00195278439543Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.Class IOpen, Classified
Z-2030-2025GE Healthcare Versana Ultrasound Systems, sold under the following names and Model/Catalog Numbers: 1) Versana Premier R3 VS, Model/Catalog Number 5938558; 2) Versana Premier R3 VA, Model/Catalog Number 5938559; 3) Versana Premier Lotus 4PP, Model/Catalog Number 5938560; 4) Versana Premier Lotus 5PP, Model/Catalog Number 5938561; 5) Versana Premier R3 VA Vet, Model/Catalog Number 5946387; 6) Versana Premier R3 VS Vet, Model/Catalog Number 5946481. 7) Versana Premier R3 Expert, Model/Catalog Number 5946950; 8) Versana Premier R3, Model/Catalog Number 5946951; 9) Versana Premier R3 Pro, Model/Catalog Number 5946952; 10) Versana Premier R3 Elite, Model/Catalog Number 5946953; 11) Versana Premier R3 Plus, Model/Catalog Number 5946954; 12) Versana Premier R3 Ultra, Model/Catalog Number 5946955; 13) Versana Premier R3 Max, Model/Catalog Number 5946956; 14) Versana Premier R3 Super, Model/Catalog Number 5946957; 15) Versana Premier R3 VS India, Model/Catalog Number 5948398; 16) Versana Premier R3 VA India, Model/Catalog Number 5948399; diagnostic ultrasound systemGE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz report. This could potentially lead to an incorrect clinical decision due to inaccuracy in the fetal size and growth calculation.Class IIOpen, Classified
Z-2029-2025GE Healthcare LOGIQ F R3 Ultrasound System, Model/Catalog Number 5943263, diagnostic ultrasound systemGE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz report. This could potentially lead to an incorrect clinical decision due to inaccuracy in the fetal size and growth calculation.Class IIOpen, Classified
Z-1638-2025GE Healthcare CARESTATION 750c A2, Model/REF Number 1012-9755-002. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.Class IOpen, Classified
Z-1637-2025GE Healthcare CARESTATION 750c A1, Model/REF Number 1012-9755-000. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.Class IOpen, Classified
Z-1636-2025GE Healthcare CARESTATION 750 A2, Model/REF Number 1012-9750-002. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.Class IOpen, Classified
Z-1635-2025GE Healthcare CARESTATION 750 A1, Model/REF Number 1012-9750-000. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.Class IOpen, Classified
Z-1634-2025GE Healthcare Carestation 650c A1, Model/REF Number 1012-9655-200. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.Class IOpen, Classified
Z-1633-2025GE Healthcare CARESTATION 650C A2, Model/REF Number 1012-9655-002. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.Class IOpen, Classified
Z-1632-2025GE Healthcare CARESTATION 650C A1, Model/REF Number 1012-9655-000. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.Class IOpen, Classified
Z-1631-2025GE Healthcare Carestation 650 A1, Model/REF Number 1012-9650-200. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.Class IOpen, Classified
Z-1630-2025GE Healthcare Carestation 650 SE, Model/REF Number 1012-9650-012. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.Class IOpen, Classified
Z-1629-2025GE Healthcare CARESTATION 650 A2, Model/REF Number 1012-9650-002. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.Class IOpen, Classified
Z-1628-2025GE Healthcare CARESTATION 650 A1, Model/REF Number 1012-9650-000. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.Class IOpen, Classified
Z-1627-2025GE Healthcare Carestation 620 A1, Model/REF Number 1012-9620-200. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.Class IOpen, Classified
Z-1626-2025GE Healthcare Carestation 620 SE, Model/REF Number 1012-9620-012. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.Class IOpen, Classified
Z-1625-2025GE Healthcare CARESTATION 620 A2, Model/REF Number 1012-9620-002. Anesthesia system.Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.Class IOpen, Classified
Z-1624-2025GE Healthcare CARESTATION 620 A1, Model/REF Number 1012-9620-000. anesthesia systemsCertain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.Class IOpen, Classified
Z-2656-2023F2-01 Frame, physiological patient monitor (with arrhythmia detection or alarms)There is a potential interruption of data communication between E-modules inserted in the F2-01 Frame and CARESCAPE ONE and CARESCAPE Canvas 1000 patient monitors if the F2-01 Frame has not been powered down within the last 120 days.Class IIOpen, Classified
Z-2956-2018GE 9100c NXT anesthesia systemsOn systems equipped with the Auxiliary Common Gas Outlet (ACGO), there is a potential for the auxiliary gas switch to unexpectedly switch from the ACGO position back to the circle position. On systems equipped with the Aux O2 + Air function the switch could unexpectedly switch from the Aux O2 + Air position back to the circle position. These situations could lead to hypoxia.Class IITerminated
Z-2955-2018GE Carestation 30 anesthesia systemsOn systems equipped with the Auxiliary Common Gas Outlet (ACGO), there is a potential for the auxiliary gas switch to unexpectedly switch from the ACGO position back to the circle position. On systems equipped with the Aux O2 + Air function the switch could unexpectedly switch from the Aux O2 + Air position back to the circle position. These situations could lead to hypoxia.Class IITerminated
Z-2954-2018GE Carestation 620/650/650c, A1 and A2 (non-US distribution) anesthesia systemsOn systems equipped with the Auxiliary Common Gas Outlet (ACGO), there is a potential for the auxiliary gas switch to unexpectedly switch from the ACGO position back to the circle position. On systems equipped with the Aux O2 + Air function the switch could unexpectedly switch from the Aux O2 + Air position back to the circle position. These situations could lead to hypoxia.Class IITerminated

Verification record

There is no verification record for this company. Most companies on this site have none: a record exists only where a company asked to be verified and MDM ran the checks, so the absence of one says nothing about the company.

Where there is a record, this is where it appears: what was checked, when it was checked, and the current status. An MDM Verified badge on another website links to a verify address on this domain that resolves to this section. A badge that does not resolve here is not ours.

What MDM checks, and what it does not

MDM coverage

MDM coverage of this company lists here, newest first. Submit news.

Is this your company?

This page is built from public FDA records. Company pages are free and factual, and stay free whether or not anyone claims them. Membership never changes editorial coverage or vetting outcomes.

Claiming lets GE Medical Systems China Co., Ltd. edit what it says about itself and list the categories it works in. It grants no badge: claiming is not verification.

Claim this page

Data as of , FDA export datePage updated Source: openFDA, public domain