GE Healthcare LOGIQ F R3 Ultrasound System, Model/Catalog Number 5943263, diagnostic ultrasound system
GE Medical Systems China Co., Ltd. began a recall of GE Healthcare LOGIQ F R3 Ultrasound System, Model/Catalog Number 5943263, diagnostic ultrasound system on . FDA classified it as Class II and gives the reason as "GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2029-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- GE Medical Systems China Co., Ltd.
- Product
- GE Healthcare LOGIQ F R3 Ultrasound System, Model/Catalog Number 5943263, diagnostic ultrasound system
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz report. This could potentially lead to an incorrect clinical decision due to inaccuracy in the fetal size and growth calculation.
- Root cause tag
- software
- FDA root cause
- Software design (manufacturing process)
- Action
- GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 05/16/2025 via letter delivered using a traceable means. The notice explained the issue, potential risk, and requested the following: "Actions to be taken by Customer/User: You may continue to use your Versana Premier R3 and LOGIQ F R3 series ultrasound systems by following the actions below: 1) If you use the Whizz report feature, reboot the system after finishing each obstetric patient examination. OR 2) Instead of using the Whizz report feature, after completing an obstetric patient examination, print all obstetric measurement and calculation results from the worksheet directly. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form to [email protected]." If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
- Quantity in commerce
- 34 units
- Distribution
- Worldwide distribution.
- Product codes and lots
- UDI/DI 00195278722430, Lot/Serial Numbers: 6310028WX0, 6310008WX0, 6310041WX0, 6310017WX0, 6310031WX0, 6310044WX0, 6310024WX0, 6310038WX0, 6310040WX0, 6310019WX0, 6310018WX0, 6310023WX0, 6310033WX0, 6310012WX0, 6310026WX0, 6310020WX0, 6310032WX0, 6310014WX0, 6310021WX0, 6310025WX0, 6310010WX0, 6310027WX0, 6310037WX0, 6310009WX0, 6310007WX0, 6310035WX0, 6310016WX0, 6310039WX0, 6310036WX0, 6310043WX0, 6310011WX0, 6310015WX0, 6310013WX0, 6310029WX0.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2029-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2030-2025 | Class II | Ongoing | Voluntary: Firm initiated |