Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR

Sysmex America, Inc.

Mundelein, IL, US

Sysmex America, Inc. appears in FDA device records in Mundelein, IL. FDA records list 31 510(k) clearances between and , no PMA records and 3 registered establishments. As of , FDA records show 16 recalls for this company, and at least one establishment registration is current.

FDA records held for this company

510(k) clearances
31
PMA records
0
Recall records
16
Registered establishments
3

510(k) clearance history

FDA records hold 31 510(k) clearances for Sysmex America, Inc. between 2003 and 2026. The busiest year on record is 2005, with 3. The most recent is 2026, with 1.

Clearances by year, as a table
510(k) clearance decisions for Sysmex America, Inc., by decision year
YearClearances
20031
20042
20053
20063
20072
20083
20091
20101
20121
20141
20151
20162
20171
20182
20191
20221
20231
20253
20261
510(k) clearances on record, most recent first
K numberDeviceProduct codeDecisionDecision date
K253212XR-Series (XR-10, XR-20) Automated Hematology AnalyzersGKZSubstantially Equivalent
K251371Sysmex XR-Series (XR-20) Automated Hematology AnalyzerGKZSubstantially Equivalent
K250943Sysmex XR-Series (XR-10) Automated Hematology AnalyzerGKZSubstantially Equivalent
K250965Automated Blood Coagulation Analyzer CN-Series (CN-6000)JPASubstantially Equivalent
K230887Sysmex XQ-Series (XQ-320) Automated Hematology AnalyzerGKZSubstantially Equivalent
K210346Sysmex XW-100 Automated Hematology AnalyzerGKZSubstantially Equivalent
K182389Sysmex XN-L Automated Hematology AnalyzerGKZSubstantially Equivalent
K182062Sysmex UD-10, Fully Automated Urine Particle Digital Imaging DeviceJOYSubstantially Equivalent
K171883Sysmex UF-5000 Fully Automated Urine Particle AnalyzerLKMSubstantially Equivalent
K172604XW-100 Automated Hematology Analyzer for CLIA Waived UseGKZSubstantially Equivalent
K160538Sysmex XN-L Automated Hematology AnalyzerGKZSubstantially Equivalent
K160541Automated Immunoassay System HISCL -800, HISCL TSH Assay Kit, HISCL TSH Calibrator, HISCL Immuno Multi ControlJLWSubstantially Equivalent
K143577XW-100 Automated Hematology Analyzer, XW QC CHECKGKZSubstantially Equivalent
K141681SYSMEX XN SERIESGKZSubstantially Equivalent
K112605SYSMEX XN-SERIESGKZSubstantially Equivalent
K091313SYSMEX MODEL XT-4000IGKZSubstantially Equivalent
K083002SYSMEX UF-500I AUTOMATED URINE PARTICLE ANALYZERLKMSubstantially Equivalent
K081890SYSMEX XT-SERIES IG PARAMETER,SYSMEX XT-SERIES RET-HE PARAMETERGKZSubstantially Equivalent
K081610SYSMEX MODEL XS-1000IC AUTOMATED HEMATOLOGY ANALYZERGKZSubstantially Equivalent
K080887SYSMEX UF-1000I, AUTOMATED URINE PARTICLE ANALYZER WITH URINALYSIS WAM SOFTWARELKMSubstantially Equivalent
K071967SYSMEX XE-5000GKZSubstantially Equivalent
K070910SYSMEX MODEL UF1000ILKMSubstantially Equivalent
K061150SYSMEX AUTOMATED HEMATOLOGY ANALYZER, XT-SERIESGKZSubstantially Equivalent
K060656SYSMEX AUTOMATED HEMATOLOGY ANALYZER MODELS XS SERIESGKZSubstantially Equivalent
K053135SYSMEX UF-100I AUTOMATED URINE CELL ANALYZERLKMSubstantially Equivalent
K051459SYSMEX MODEL XE-2100DC AUTOMATED HEMATOLOGY ANALYZERGKZSubstantially Equivalent
K051199IPF (IMMATURE PLATELET FRACTION) PARAMETER ON THE SYSMEX MODEL XE-2100 AUTOMATED HEMATOLOGY ANALYZERGKZSubstantially Equivalent
K050589RET-HE PARAMETER ON THE SYSMEX MODEL XE-2100 AUTOMATED HEMATOLOGY ANALYZERGKZSubstantially Equivalent
K040073SYSMEX AUTOMATED HEMATOLOGY ANALYZER, XE-2100 SERIES (XE-2100, XE-2100L)GKZSubstantially Equivalent
K032677SYSMEX POCH-100IGKLSubstantially Equivalent
K032039IMMATURE GRANULOCYTE (IG) PARAMETER ON THE SYSMEX MODEL XE-2100 AUTOMATED HEMATOLOGY ANALYZERGKZSubstantially Equivalent

Registered establishments

FDA registered establishments for Sysmex America, Inc., with the product codes listed at each
RegistrationFEIEstablishmentAddressRegistration expiresProduct codes listed
14226811000515253SYSMEX AMERICA, INC.ONE SYSMEX WAY, Mundelein, IL 60060 US2026
30094956213009495621Sysmex America, Inc.SAI - Multi-Purpose Building, Buffalo Grove, IL 60089 US2026
30097114783009711478Sysmex America, Inc.577 Aptakisic Rd, Lincolnshire, IL 60069 US2026and 20 more, listed in full below

Registration numbers link to the FDA registration and listing record for that establishment.

FDA product codes

Recall records

Device recall records naming this company, most recent first
RecallProductReasonClassificationInitiatedStatus
Z-1136-2026TS-10/TS-10H Tube Sorter, automatic sorting system for sorting barcoded sample tubes, Catalog Number BN168792, running software versions 00-17Under specific conditions, the lift's electromagnetic holding force may disengage unexpectedly, allowing unintended downward movement of the lift.Class IIOpen, Classified
Z-1113-2025Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341. Product usage for automated pipetting, diluting, and specimen processing workstations for flow cytometric analysis.Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.Class IIOpen, Classified
Z-2687-2020Sysmex PS-10 Sample Preparation System Catalog number:BQ716341 - Product Usage: is a fully automated, user configurable, open system intended for use to prepare human specimens for subsequent analysis on flow cytometers.Insufficient amount of antibody without an error message or alarmClass IITerminated
Z-1703-2020The CF-70 instrument (product code: LXG; Regulation Number: 862.2050) receives magazines containing stained blood smears from SP-50, selects only smears that require analysis, and transports them to the DI-60 for further analysis. After completing the analysis on DI-60 the smears are placed back in magazines.Software mismatch-When the software versions between the SP-50 and CF-70 are not matched, and an error condition occurs requiring the operator to open the cover door of the CF-70 to remove slides or slide magazines from the CF-70, the operation of the CF-70 does not halt.Class IITerminated
Z-1717-2020Sysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10 analyzers.When excessive amounts of urine spills on the rack transport track and reaches the electrical components of the CV-11 conveyor, there is a potential for smoke and/or sparks which may present a combustion hazard.Class IITerminated
Z-2953-2018Sysmex XN-Series Fluorocell PLT, Catalog #CD994563. Distributed in the U.S.A. by Sysmex America, Inc., Lincolnshire, IL. Fluorocell PLT is used to stain platelets in a diluted blood sample for the counting of platelets with the Sysmex XN-Series automated hematology analyzers.The fluorescent platelet counts (PLT-F) and immature platelet fraction (IPF) values can be falsely low when using affected lots of Fluorocell PLT.Class IITerminated
Z-0045-2018Sysmex Fluorocell(R) PLT Fluorocell PLT is used to stain platelets in a diluted blood sample for the counting of platelets with the Sysmex XN-Series automated hematology analyzersThere is potential that fluorescent platelet (PLT-F) and immature platelet fraction (IPF%, #) results obtained using the lots listed above may exhibit lower than actual PLT-F and IPF values. The issue may be recognized by a large discrepancy between the impedance platelet (PLT-I) result and the PLT-F result. In most cases, the falsely decreased PLT-F results display a PLT Abn Scattergram message with an asterisk beside the result or an Action Message: Difference between PLT and PLT- F. Check the results. These are indicators of unreliable data. In rare instances, platelet flags may be absent. The PLT-I results are unaffected.Class IICompleted
Z-2034-2017Sysmex, Automated Hematology Analyzer XN Series, Sysmex Corporation, 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073 JapanReported customer complaints of automated hematology analyzers catching fire due to misalignment of an internal protective sheet that shields electronic components from salty reagents. Build-up of salt deposits on unprotected electronic components may combust due to heat generated by electronics when instrument cycles. Completed CAPA identified the root cause as supplier providing defective component (sensor ready valve) and the misalignment of the electronic protective shield during manufacturing of the device.Class IITerminated
Z-2770-2016Fluorocell PLT reagent, Catalog #CD994563 Fluorocell PLT is used to stain platelets in a diluted blood sample for the counting of platelets with the Sysmex XN-Series automated hematology analyzersThere is a potential that fluorescent platelet (PLT-F) and immature platelet fraction (IPF%) results performed using the recalled lots may exhibit lower than expected PLT-F and IPF values. The issue may be recognized by a large discrepancy between the Impedance platelet (PLT-I) value and a false low PLT-F value. In most cases, the falsely decreased PLT-F results displayed a "PLT Abn Scattergram" message with asterisks beside the results indicating the data is unreliable. In rare instances, platelet flags may be absent. The PLT-I counts are unaffected. The PLT-F is a reflex test used to confirm PLT-I results due to flagging or low values. As such, the likely impact to patient results is low. Affected results display a discrepancy between the PLT-I and PLT-F, with lower than expected PLT-F results that do not match the patient's clinical picture.Class IITerminated
Z-0100-2014pocH-100i; sysmex; Automated Hematology Analyzer; 7.3" x 13.8" x 18.1" / 30.8 lbs.; Sysmex Corporation; 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073, Japan; Sysmex America, Inc; One Nelson C. White Pkwy, Mundeline, IL 60060, U.S.A. The Sysmex pocH-100i Automated Hematology Analyzer is an automated cell counter intended for in vitro diagnostic use in CLIA non-waived clinical laboratories (not for Point of Care use in a CLIA waived laboratory). The Complete Blood Cel (CBC) count test parameters include: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, and NEUT%/#, LYMPH %/#, MIXED %/# (EO, BSO, MONO).The data printed on the thermal printer may omit a digit or decimal point in the results or sample identification number.Class IITerminated
Z-0949-2014The UF-1000i is a fully automated urine particle analyzer intended for vitro diagnostic use in urinalysis. The instrument can perform screening for abnormal samples with a high degree of accuracy, so it increases automation and efficiency in laboratories.The current Century Break patch did not contain the same setting for the Century Break year for both the Red Hat Linux and GUI files. Due to the mismatch in file settings, the software could not properly calculate a purge date for the database files based on the customer data retention setting in the application. This resulted in the database files meeting its maximum capacity and causing system pClass IITerminated
Z-0256-2013Sysmex XE-2100C Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe, Japan; Sysmex America, One Nelson C. White Parkway, Mundelein, IL 60060; catalog #053-2311-4 The Sysmex XE-2100C is an automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories.There is a variation of reticulocyte counts between XE Series (XE-2100, XE-5000) and XT-Series (XT-2000i, XT-4000i) hematology analyzers that may result in a retic counts that are an average of 21% lower on the XE-Series because of a low bias.Class IITerminated
Z-0255-2013Sysmex XE-5000 Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe, Japan; Sysmex America, One Nelson C. White Parkway, Mundelein, IL 60060; catalog #063-7591-7 The Sysmex XE-5000 is an automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XE-5000 performs analyses using the following methods: RF/DC Detection Method, Sheath Row DC Detection Method, and Flow Cytometry Methods using a Semiconductor Laser. Particle characterization and identification is based on detection of forward scatter, fluorescence and adaptive cluster analysis. Using the same reagents as the XE-2100, the XE-5000 automatically classifies cells from blood and body fluids and carries out all processes automatically from aspiration of the sample to outputting the results.There is a variation of reticulocyte counts between XE Series (XE-2100, XE-5000) and XT-Series (XT-2000i, XT-4000i) hematology analyzers that may result in a retic counts that are an average of 21% lower on the XE-Series because of a low bias.Class IITerminated
Z-0254-2013Sysmex XE-2100 Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe, Japan; Sysmex America, One Nelson C. White Parkway, Mundelein, IL 60060; catalog #983-1341-1 The Sysmex XE-2100 is an automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XE-2100 performs hematology analyses according to the RF/DC detection method, HydroDynamic Focusing (DC Detection), flow cytometry method (using a semiconductor laser), and SLS-hemoglobin method.There is a variation of reticulocyte counts between XE Series (XE-2100, XE-5000) and XT-Series (XT-2000i, XT-4000i) hematology analyzers that may result in a retic counts that are an average of 21% lower on the XE-Series because of a low bias.Class IITerminated
Z-3063-2011Sysmex Stromatolyser-IM; a lytic reagent for use with the XE-2100, XE-2100L and XE-5000 Series Sysmex Automated Hematology Analyzers; the reagent solution is supplied in a 10-liter cubetainer (cardboard cube with a 10-liter plastic liner); Sysmex America, Inc., Mundelein, IL 60060; catalog number SIM-220A. The STROMATOLYSER-IM lysing reagent is used to destroy red blood cells to aid in the counting and characterization on immature white blood cells. This reagent is used on the Sysmex XE-Series and SE Series hematology analyzers.The affected lots of Stromatolyser-IM lysing reagent were found to be contaminated with Pseudomonas bacteria, which resulted in rising IMI channel background counts, increased false "Platelet Clump" flags and a strong sulfur-like odor when used on the XE-2100, XE-2100L and XE-5000 analyzers.Terminated
Z-2797-2011Sysmex Stromatolyser-IM; a lytic reagent for use with the XE-2100, XE-2100L and XE-5000 Series Sysmex Automated Hematology Analyzers; the reagent solution is supplied in a 10-liter cubetainer (cardboard cube with a 10-liter plastic liner); Sysmex America, Inc., Mundelein, IL 60060; catalog number SIM-220A. The product is a lytic reagent used to lyse (destroy) red blood cells for hemoglobin determinations or aid in the counting of white blood cells.The affected lots of Stromatolyser-IM lysing reagent was found to be contaminated with Pseudomonas bacteria, which resulted in rising IMI channel background counts, increased false "Platelet Clump" flags and a strong sulfur-like odor when used on the XE-2100, XE-2100L and XE-5000 analyzers.Terminated

Verification record

There is no verification record for this company. Most companies on this site have none: a record exists only where a company asked to be verified and MDM ran the checks, so the absence of one says nothing about the company.

Where there is a record, this is where it appears: what was checked, when it was checked, and the current status. An MDM Verified badge on another website links to a verify address on this domain that resolves to this section. A badge that does not resolve here is not ours.

What MDM checks, and what it does not

MDM coverage

MDM coverage of this company lists here, newest first. Submit news.

Is this your company?

This page is built from public FDA records. Company pages are free and factual, and stay free whether or not anyone claims them. Membership never changes editorial coverage or vetting outcomes.

Claiming lets Sysmex America, Inc. edit what it says about itself and list the categories it works in. It grants no badge: claiming is not verification.

Claim this page

Data as of , FDA export datePage updated Source: openFDA, public domain