Wipro GE Healthcare Private Ltd.

Bangalore, IN

WIPRO GE HEALTHCARE PRIVATE LTD appears in FDA device records in Bangalore, IN. FDA records list 5 510(k) clearances between and , no PMA records and 3 registered establishments. As of , FDA records show 10 recalls for this company, and at least one establishment registration is current.

FDA records held for this company

510(k) clearances
5
PMA records
0
Recall records
10
Registered establishments
3

510(k) clearance history

FDA records hold 5 510(k) clearances for Wipro GE Healthcare Private Ltd. between 2010 and 2013. The busiest year on record is 2012, with 2. The most recent is 2013, with 1.

Clearances by year, as a table
510(k) clearance decisions for Wipro GE Healthcare Private Ltd., by decision year
YearClearances
20101
20111
20122
20131
510(k) clearances on record, most recent first
K numberDeviceProduct codeDecisionDecision date
K1236811.5T 4CH LARGE FLEX COIL, 1.5 4CH SMALL FLEX COILMOSSubstantially Equivalent
K121625LULLABY TM WARMERFMTSubstantially Equivalent
K120168LULLABY LED PHOTTHERAPY SYSTEMLBISubstantially Equivalent
K110556MINI TELEMETRY SYSTEMHGMSubstantially Equivalent
K102104GE VIVID P3 DIAGNOSTIC ULTRASOUND IMAGING SYSTEMIYNSubstantially Equivalent

Registered establishments

FDA registered establishments for Wipro GE Healthcare Private Ltd., with the product codes listed at each
RegistrationFEIEstablishmentAddressRegistration expiresProduct codes listed
96139361000571342WIPRO GE HEALTHCARE PRIVATE LTD122, Part I, EXPORT PROMOTIONAL INDUSTRIAL PARK, WHITEFIELD, BANGALORE Karnataka 560 066 IN2026
30077348883007734888WIPRO GE HEALTHCARE PRIVATE Ltd.122 (Part 1), JOHN F. WELCH TECHNOLOGY CENTRE, ODYSSEY AND GALILEO, BANGALORE Karnataka 560066 IN2026and 18 more, listed in full below
96172773003190524Wipro GE Healthcare Private Ltd.4, Kadugodi Industrial Area, Bangalore Karnataka 560067 IN2026and 8 more, listed in full below

Registration numbers link to the FDA registration and listing record for that establishment.

FDA product codes

Recall records

Device recall records naming this company, most recent first
RecallProductReasonClassificationInitiatedStatus
Z-0214-2026GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5, Part Number 5867474-03; Medical Display, cathode ray tubeCarescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.Class IIOpen, Classified
Z-0213-2026GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04; Medical Display, cathode ray tubeCarescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.Class IIOpen, Classified
Z-3081-2024GE Panda iRES Warmer, Model/Catalog Numbers: 1) M1112198-01045518; 2) M1112198-01045579; 3) M1112198-01047148; 4) M1112198-01047271; 5) M1112198-01048731; 6) M1112198-01048740; 7) M1112198-01052386; 8) M1112198-01054697; 9) M1112198-01068942; 10) M1112198-01084972; 11) M1112198-01090848; 12) M1112198-01093457; 13) M1112198-01117766; 14) M1112198-01139893; 15) M1112198-01156745; 16) M1112198-01167230; 17) M1112198-01177090; 18) M1112198-01183599; 19) M1112198-01185372; 20) M1112198-01193801; 21) M1112198-01200764; 22) M1112198-01245548; 23) M1112198-01253742; 24) M1112198-01274059. Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment.GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design. However, because the electrical safety testing in manufacturing was incomplete for these devices, there is the potential for leakage current to exceed IEC 60601 limits.Class IIOpen, Classified
Z-3080-2024GE Giraffe Warmer, Model/Catalog Number M1118179-01090979. Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment.GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design. However, because the electrical safety testing in manufacturing was incomplete for these devices, there is the potential for leakage current to exceed IEC 60601 limits.Class IIOpen, Classified
Z-0962-2024GE Healthcare Lullaby Incubator Model Hood Door FRU kit, Part Numbers: M1169767, M1160674, M1187512, M1206646, M1166822GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.Class IOpen, Classified
Z-0961-2024GE Healthcare Lullaby Incubator Model Hood FRU kit, Part Numbers: M1169566, M1160659, M1206259, M1206260, M1166836, M1166814GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.Class IOpen, Classified
Z-0960-2024GE Healthcare Care Plus Hood Door FRU kit, Part Numbers: 6600-0082-850, 6600-0504-800, 6600-0148-800, and 6600-0970-500GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.Class IOpen, Classified
Z-0959-2024GE Healthcare Care Plus Hood Field Replacement Unit (FRU) kit, Part Numbers: 6600-0047-851, 6600-0033-850, 6600-0081-850, 6600-0961-500, 6600-0083-850GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.Class IOpen, Classified
Z-0958-2024GE Healthcare neonatal incubator, models: a) Lullaby Incubator, b) Lullaby Incubator XP, c) Lullaby Incubator TRGE HealthCare has become aware that the bedside panel of Care Plus, Care Plus Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.Class IOpen, Classified
Z-0957-2024GE Healthcare neonatal incubator, models: a) Care Plus, b) Care Plus 1000, c) Care Plus 2000, d) Care Plus 3000, e) Care Plus 4000, Plus 4000GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.Class IOpen, Classified

Verification record

There is no verification record for this company. Most companies on this site have none: a record exists only where a company asked to be verified and MDM ran the checks, so the absence of one says nothing about the company.

Where there is a record, this is where it appears: what was checked, when it was checked, and the current status. An MDM Verified badge on another website links to a verify address on this domain that resolves to this section. A badge that does not resolve here is not ours.

What MDM checks, and what it does not

MDM coverage

MDM coverage of this company lists here, newest first. Submit news.

Is this your company?

This page is built from public FDA records. Company pages are free and factual, and stay free whether or not anyone claims them. Membership never changes editorial coverage or vetting outcomes.

Claiming lets Wipro GE Healthcare Private Ltd. edit what it says about itself and list the categories it works in. It grants no badge: claiming is not verification.

Claim this page

Data as of , FDA export datePage updated Source: openFDA, public domain