GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number...

Wipro GE Healthcare Private Ltd. / initiated 2025-09-05 / Open, Classified / root cause: software

Wipro GE Healthcare Private Ltd. began a recall of GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04; Medical Display, cathode ray tube on . FDA classified it as Class II and gives the reason as "Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0213-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Wipro GE Healthcare Private Ltd.
Product
GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04; Medical Display, cathode ray tube
Product code
DXJ Display, Cathode-Ray Tube, Medical
Reason for recall
Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.
Root cause tag
software
FDA root cause
Software design
Action
GE HealthCare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 09/05/2025, by letter delivered via traceable means. The notice explained the issues and requested the following: "Actions to be taken by Customer/User Pending correction by GE HealthCare, you can continue to use your device by following the instructions below. 1. Determine the number of central stations that are connected to the network including those in any offsite Central Monitoring Unit. 2. Ensure the number of connected central stations are fewer than 118 or more than 120. If you need assistance completing steps 1 or 2, please contact GE HealthCare Service or your local Service Representative. Note: Any versions prior to CSCS V3.0.5 are not affected by this issue. Ensure all users are aware of, understand, and follow these instructions. Please retain this document for your records. GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Quantity in commerce
173 units
Distribution
Worldwide distribution.
Product codes and lots
GTIN 00195278343512, Serial Numbers: STU25210418SA, STU25210417SA, STU25210419SA, STU25250440SA, STU25250439SA, STU25260447SA, STU25210407SA, STU25210406SA, STU24200003SA, STU24200006SA, STU24200005SA, STU24200004SA, STU23240015SA, STU24240019SA, STU24240020SA, STU24240007SA, STU24240022SA, STU24240021SA, STU24330020SA, STU24240014SA, STU24240013SA, STU24330019SA, STU24330022SA, STU24330016SA, STU24240005SA, STU24240008SA, STU24240012SA, STU24330018SA, STU24330021SA, STU24330017SA, STU23160011SA, STU23240007SA, STU24310001SA, STU25210433SA, STU25210432SA, STU25200390SA, STU25200392SA, STU25200391SA, STU25200399SA, STU25200397SA, STU25200395SA, STU25200393SA, STU25200398SA, STU25200396SA, STU25250438SA, STU25270475SA, STU25270456SA, STU25270468SA, STU25270458SA, STU25270466SA, STU25270476SA, STU25270469SA, STU25270473SA, STU25270459SA, STU25270474SA, STU25270467SA, STU25210430SA, STU25210429SA, STU25180343SA, STU25190347SA, STU25190348SA, STU24200007SA, STU25190360SA, STU25240436SA, STU25190361SA, STU25190362SA, STU25190363SA, STU25240437SA, STU25190359SA, STU24510027SA, STU24510032SA, STU24510010SA, STU24510020SA, STU24510017SA, STU24510026SA, STU24510018SA, STU24510024SA, STU24510025SA, STU24510031SA, STU24510011SA, STU24510029SA, STU24510023SA, STU24510030SA, STU24510021SA, STU24510019SA, STU24510022SA, STU25190350SA, STU25200388SA, STU25200389SA, STU25210412SA, STU25190349SA, STU25190381SA, STU25190376SA, STU25190383SA, STU25190374SA, STU25190367SA, STU25190372SA, STU25190379SA, STU25190368SA, STU25190352SA, STU25190358SA, STU25190370SA, STU25190366SA, STU25190355SA, STU25190354SA, STU25190382SA, STU25190380SA, STU25190371SA, STU25190377SA, STU25190353SA, STU25190351SA, STU25190364SA, STU25190357SA, STU25190384SA, STU25190369SA, STU25190373SA, STU25190356SA, STU25190365SA, STU25190378SA, STU25190385SA, STU25210410SA, STU25210411SA, STU25210409SA, STU25230435SA, STU25180346SA, STU25180344SA, STU25180345SA, STU25210413SA, STU25210414SA, STU25210423SA, STU25210431SA, STU25210415SA, STU25270453SA, STU25260450SA, STU25260451SA, STU25290488SA, STU25210408SA, STU25210400SA, STU25270454SA, STU25210404SA, STU25210421SA, STU25270464SA, STU25210420SA, STU25270462SA, STU25210425SA, STU25270461SA, STU25210426SA, STU25210428SA, STU25210427SA, STU25270463SA, STU25270460SA, STU24460058SA, STU24460059SA, STU24460071SA, STU24460060SA, STU25200387SA, STU25210401SA, STU25200386SA, STU25210424SA, STU25270457SA, STU25270465SA, STU25300492SA, STU25210402SA, STU25210403SA, STU25250446SA, STU25210422SA, STU25210405SA, STU25260449SA, STU25290481SA, STU25290482SA, STU25290483SA, STU25290484SA, STU25210416SA.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0213-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0213-2026Class IIOngoingVoluntary: Firm initiated
Z-0214-2026Class IIOngoingVoluntary: Firm initiated