Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0831-2012

Surgical Headlight labeled in part *Gulf Medical Fiberoptics*OLDSMAR, FLORIDA U.S.A.* Model Numbers...

Gulf Fiberoptics, Inc. / initiated 2011-05-13 / Terminated / root cause: manufacturing process

Gulf Medical Fiberoptics, Inc. began a recall of Surgical Headlight labeled in part *Gulf Medical Fiberoptics*OLDSMAR, FLORIDA U.S.A.* Model Numbers: GMF-H001, GMF-H100, GMF-H101, GMF-H102, GMF-H103... on . The reason FDA records is "Gulf Medical Fiberoptics, Inc. recalled their Surgical Headlight because the cleaning instructions for the exterior surfaces of the headlight , headgear and fiber cable with 70% isopropyl alcohol had not been validated". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0831-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Gulf Fiberoptics, Inc.
Product
Surgical Headlight labeled in part *Gulf Medical Fiberoptics*OLDSMAR, FLORIDA U.S.A.* Model Numbers: GMF-H001, GMF-H100, GMF-H101, GMF-H102, GMF-H103, GMF-H104, GMF-H105, and GMF-H106. The Surgical Headlight is a passive luminaire. It is illuminated by a source of light by way of the fiber optics and provides supplemental lighting for medical and surgical procedures.
Product code
FST Light, Surgical, Fiberoptic
Reason for recall
Gulf Medical Fiberoptics, Inc. recalled their Surgical Headlight because the cleaning instructions for the exterior surfaces of the headlight , headgear and fiber cable with 70% isopropyl alcohol had not been validated.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On May 16, 2011, Gulf Medical Fiberoptics, Inc. issued to all customers an updated letter to their May 13, 2011 "Field Correction of Instructions For Use" informing them to discontinue the use of 70% isopropyl to clean the surfaces of the device. Enclosed with the letter was an updated Surgical Headlight User Guide with instructions to discard/shred/recycle the prior guide. The firm asked customers to return a form acknowledging the receipt and action from the letter. Customers with questions should contact Customer Service department manager Jacqueline Armand at 813-855-6618 ext. 427.
Quantity in commerce
381 units
Distribution
Worldwide Distribution - USA, including the states of FL, WI, WA, AL, CA, MO, MI, OH, VA, CO, TX, TN, NC and the countries of Turkey, Brazil, Belgium, South Korea, Panama, Mexico, Ireland, Taiwan (ROC), Italy, Argentina, India, Spain, Germany, and Venezuela.
Product codes and lots
Model Numbers: GMF-H001, GMF-H100, GMF-H101, GMF-H102, GMF-H103, GMF-H104, GMF-H105, GMF-H106.