Z-3221-2011
TSX-101: Aquilion 32/64; System, X-Ray, Tomography, Computed
Canon Medical Systems USA, Inc. / initiated 2010-12-08 / Terminated / root cause: software
Toshiba American Medical Systems Inc began a recall of TSX-101: Aquilion 32/64; System, X-Ray, Tomography, Computed on . The reason FDA records is "There is the potential that the current may not be controlled correctly in examinations using the ECG modulation function, which adjusts the x-ray output in synchronization with the ECG waveform". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-3221-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- TSX-101: Aquilion 32/64; System, X-Ray, Tomography, Computed.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- There is the potential that the current may not be controlled correctly in examinations using the ECG modulation function, which adjusts the x-ray output in synchronization with the ECG waveform.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Toshiba America Medical Systems (TAMS) issued an Urgent: Medical Device Correction notification to consignees via USPS return receipt mail. The letter identified the affected product, the problem, and corrective action. The firm plans to install revised firmware to prevent the occurrence of the problem. A Toshiba Service Representative will contact customers to schedule an appointment for installation. The firm asked customers to stop using the Power Reduction option during ECG modulated scanning until the new firmware has been installed. Customers are to share the information in the letter with all users, reviewing radiologist, and the clinical engineering or Biomedical group at their facility. Customers are also to complete and return the attached reply form via fax or e-mail to TAMS as proof of notification. Questions pertaining to the letter should be directed to 800-421-1968 or customers' local Toshiba Representative at 800-521-1968.
- Quantity in commerce
- 528 units
- Distribution
- Nationwide Distribution.
- Product codes and lots
- Units manufactured from January 2005 to September 2007