Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR

Canon Medical Systems USA, Inc.

Tustin, CA, US

Canon Medical Systems USA, Inc. appears in FDA device records in Tustin, CA. FDA records list one 510(k) clearance, no PMA records and one registered establishment. As of , FDA records show 176 recalls for this company, and at least one establishment registration is current.

FDA records held for this company

510(k) clearances
1
PMA records
0
Recall records
176
Registered establishments
1

FDA records under these names

3 names

FDA records resolved to Canon Medical Systems USA, Inc. carry these names as well, as FDA wrote them. Each is the firm named on a registration, a 510(k), a PMA, a recall or an enforcement report that was joined to this company on a shared FDA identifier or on a matching name.

  • Canon Medical System, USA, INC.enforcement report, recall
  • Toshiba American Med Sys Increcall
  • Toshiba American Medical Systems Incenforcement report, recall

510(k) clearance history

FDA records hold 1 510(k) clearance for Canon Medical Systems USA, Inc. in 2018.

Clearances by year, as a table
510(k) clearance decisions for Canon Medical Systems USA, Inc., by decision year
YearClearances
20181
510(k) clearances on record, most recent first
K numberDeviceProduct codeDecisionDecision date
K182223Aquilion ONE Self-Propelled Scan Base Kit for IVR-CT, CGBA-034AJAKSubstantially Equivalent

Registered establishments

FDA registered establishments for Canon Medical Systems USA, Inc., with the product codes listed at each
RegistrationFEIEstablishmentAddressRegistration expiresProduct codes listed
20205632020563Canon Medical Systems USA, Inc.2441 Michelle Dr, TUSTIN, CA 92780 US2026No device listings on this registration

Registration numbers link to the FDA registration and listing record for that establishment.

FDA product codes

Recall records

Showing 25 of 176
Device recall records naming this company, most recent first
RecallProductReasonClassificationInitiatedStatus
Z-3014-2026Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.Due to the fixing screws of ceiling movement gear in ceiling suspended arm that are part of the X-ray interventional systems may become loose.Class IIOpen, Classified
Z-1018-2026VANTAGE TITAN 3T Model MRT-3010/MEXL-3010There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.Class IIOpen, Classified
Z-1017-2026VANTAGE GALAN 3T Model MRT-3020/MEXL-3020There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.Class IIOpen, Classified
Z-1179-2026Alphenix INFX-8000V, interventional fluoroscopic x-ray systemIt has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayedClass IIOpen, Classified
Z-1178-2026Alphenix INFX-8000H, interventional fluoroscopic x-ray systemIt has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayedClass IIOpen, Classified
Z-1177-2026Alphenix INFX-8000F, interventional fluoroscopic x-ray systemIt has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayedClass IIOpen, Classified
Z-1176-2026Alphenix INFX-8000C, interventional fluoroscopic x-ray systemIt has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayedClass IIOpen, Classified
Z-0327-2026VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.Class IIOpen, Classified
Z-0037-2026This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.It has been found that when SPOT Fluoro is first turned on, the default ROI is selected and the collimator blades move to the boundary of the ROI, and the static image is overlayed around the ROI, covering the collimated area. If the customer changes the ROI, the collimator blades and static images are adjusted accordingly. However, when the acquisition program is changed, the collimator returns to the default ROI position which may be larger than the specified one, leaving the X-ray area partially covered by the static image. This is a software problem with the SPOT Fluoro program and will not occur if the function is not used.Class IIOpen, Classified
Z-0284-2025Cartesion Prime, Digital PET-CT, Model: PCD-1000A/3D.004, PCD-1000A/3D, is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.When PET-CT system is executing reconstruction, if PET acquisition for another patient is performed (or PET reconstruction for another patient is performed from raw data processing), PET reconstruction control processing may hang up and PET reconstruction may stop as a result, which may necessitate repeat PET/CT examinations.Class IIOpen, Classified
Z-2041-2024MRI systems: Vantage Orian MRT-1550 (MEXL-1550), Vantage Elan MRT-2020 (MEXL-1520), Vantage Titan MRT-1504 (MEXL-1504), Vantage Titan 3T MRT-3010 (MEXL-3010), Vantage Galan 3T MRT-3020 (MEXL-3020)The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.Class IIOpen, Classified
Z-1250-2024Vantage Titan, Model: MRT-1504For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.Class IIOpen, Classified
Z-1249-2024MRI system: Vantage Elan, Model: MRT-2020For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.Class IIOpen, Classified
Z-1248-2024MRI system: Vantage Orian, Vantage Fortian, Model: MRT-1550For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.Class IIOpen, Classified
Z-1155-2024The INFX-8000F is designed to take advantage of the latest technological innovations to reduce dose and to save time. A revolutionary graphic user interface and a multi-tasking computer enable the system to fully meet the requirement for optimum image quality, safety, ease of use, advanced efficiency and improved workflow. The advanced floormounted system INFX-8000F uses a new flat panel detector to present fluoroscopic and fluorographic images of high quality.VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.Class IIOpen, Classified
Z-1154-2024The device is an X-ray system that is capable of radiographic and fluoroscopic studies and is used in an interventional setting. The system consists of a C-arm which is equipped with an X-ray tube, beam limiter and X-ray receptor, X-ray controller, computers with system and processing software, and a patient radiographic table.VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.Class IIOpen, Classified
Z-2392-2023INFX-8000C XGCP-930AA TABLE SIDE CONSOLE, INFX-8000C XGCP-930BA TABLE SIDE CONSOLEThe C-arm, used to set the position and angle of the x-ray exposure field, may not function due to the tableside console PWB, which needs replacement. There may be a delay in the procedure, but procedures can continue using the operators console, the switches on the back of the column, switches on the side of the flat panel detector, and the emergency stop will continue to function normally.Class IIOpen, Classified
Z-0198-2023Canon Aplio ultrasound system, Model: Aplio i900 TUS-AI900, Aplio i800 TUS-AI800, Aplio i700 TUS-AI700, Aplio a550 CUS-AA550, Aplio a450 CUS-AA450After completion, ultrasound system's Stress Echo may not terminate if user does not manually terminate the protocol. During next patient exam, an abnormal confirmation message will be displayed "Same view was saved in this phase. Do you use this image with StressEcho Review?" If "no", acquired images not saved, and if "yes" images assigned to the previous patient; could cause treatment delay.Class IIOpen, Classified
Z-1289-2022System: Alphenix 4D CT in combination with CAS-930A (C arm Support) Model Number: INFX-8000C/BF, INFX-8000C/BH, INFX-8000C/SWCT operation may be restricted by an interlock which is a result of an system error and the CT system may stop operating properly after an attempt to cancel the error message has been initiated.Class IIOpen, Classified
Z-1666-2022INFX-8000V. For radiographic and fluoroscopic studies and intervention.The fluoroscopic dose rate might exceed the conforming value during biplane fluoroscopy.Class IIOpen, Classified
Z-0492-2022Flat Panel Detector, Model Number: TFP-800A, used with the Alphenix Systems, Item Numbers: INFX-8000V, INFX-8000C, INFX-8000F, and INFX-8000H.Abnormal images such as horizontal stripes may be displayed and recovery may not not be possible by restarting the system. The worst case scenario is an operator who is working without a workable image may remove a catheter from the patient's blood vessel and damage the blood vessel.Class IIOpen, Classified
Z-2047-2021Canon Vantage Galan 3T, REF: MRT-3020 - Product Usage: use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.The MRI system contains a component that does not meet the appropriate flame retardancy-rating which could result in smoke or fire.Class IIOpen, Classified
Z-2046-2021Canon Vantage Titan 3T, Premium Open-Bore MRI System, REF: MRT-3010 - Product Usage: use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.The MRI system contains a component that does not meet the appropriate flame retardancy-rating which could result in smoke or fire.Class IIOpen, Classified
Z-2045-2021Canon Vantage Tiitan 1.5T, Premium Open-Bore MRI System, REF: MRT-1510 - Product Usage: use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.The MRI system contains a component that does not meet the appropriate flame retardancy-rating which could result in smoke or fire.Class IIOpen, Classified
Z-2044-2021Toshiba Vantage Tiitan 1.5T with Standard gradient, Premium Open-Bore MRI System, REF: MRT-1504 - Product Usage: use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.The MRI system contains a component that does not meet the appropriate flame retardancy-rating which could result in smoke or fire.Class IIOpen, Classified

Verification record

There is no verification record for this company. Most companies on this site have none: a record exists only where a company asked to be verified and MDM ran the checks, so the absence of one says nothing about the company.

Where there is a record, this is where it appears: what was checked, when it was checked, and the current status. An MDM Verified badge on another website links to a verify address on this domain that resolves to this section. A badge that does not resolve here is not ours.

What MDM checks, and what it does not

MDM coverage

MDM coverage of this company lists here, newest first. Submit news.

Is this your company?

This page is built from public FDA records. Company pages are free and factual, and stay free whether or not anyone claims them. Membership never changes editorial coverage or vetting outcomes.

Claiming lets Canon Medical Systems USA, Inc. edit what it says about itself and list the categories it works in. It grants no badge: claiming is not verification.

Claim this page

Data as of , FDA export datePage updated Source: openFDA, public domain