Anesthesia Machines (digital display assembly); A gas machine for anesthesia is a device used to administer...
Penlon, Ltd. began a recall of Anesthesia Machines (digital display assembly); A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a... on . The reason FDA records is "Penlon, Ltd has identified the potential for the premature failure of a component in some Electronic Flowmeter sensor display unit PCB assemblies". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1318-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Penlon, Ltd.
- Product
- Anesthesia Machines (digital display assembly); A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation.
- Product code
- BSZ Gas-Machine, Anesthesia
- Reason for recall
- Penlon, Ltd has identified the potential for the premature failure of a component in some Electronic Flowmeter sensor display unit PCB assemblies.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Penlon Limited sent a Technical Bulletin dated November 22, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. As a precaution a field upgrade part was released that required all existing PCB assemblies to be removed and the updated item installed. The old PCB must be returned to Penlon with the serial number of the affected machine. The required replacement parts were included with the technical bulletin. For further information customers should contact the Technical Support Team at [email protected].
- Quantity in commerce
- 160 units
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Peru, Uganda, Kenya, Malaysia, Poland, Saudi Arabia, United Arab Emirates, Philippines, Equador, and Thailand
- Product codes and lots
- None