Z-1318-2012

Anesthesia Machines (digital display assembly); A gas machine for anesthesia is a device used to administer...

Penlon, Ltd. / initiated 2011-11-29 / Terminated / root cause: design

Penlon, Ltd. began a recall of Anesthesia Machines (digital display assembly); A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a... on . The reason FDA records is "Penlon, Ltd has identified the potential for the premature failure of a component in some Electronic Flowmeter sensor display unit PCB assemblies". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1318-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Penlon, Ltd.
Product
Anesthesia Machines (digital display assembly); A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation.
Product code
BSZ Gas-Machine, Anesthesia
Reason for recall
Penlon, Ltd has identified the potential for the premature failure of a component in some Electronic Flowmeter sensor display unit PCB assemblies.
Root cause tag
design
FDA root cause
Component design/selection
Action
Penlon Limited sent a Technical Bulletin dated November 22, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. As a precaution a field upgrade part was released that required all existing PCB assemblies to be removed and the updated item installed. The old PCB must be returned to Penlon with the serial number of the affected machine. The required replacement parts were included with the technical bulletin. For further information customers should contact the Technical Support Team at [email protected].
Quantity in commerce
160 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Peru, Uganda, Kenya, Malaysia, Poland, Saudi Arabia, United Arab Emirates, Philippines, Equador, and Thailand
Product codes and lots
None