Z-1951-2012Class II

AEM Disposable Electrodes, Device, Electrosurgical, Cutting and Coagulation and Accessories. AEM Disposable...

Encision, Inc. / initiated 2012-06-15 / Terminated / root cause: design

Encision, Inc. began a recall of AEM Disposable Electrodes, Device, Electrosurgical, Cutting and Coagulation and Accessories. AEM Disposable Electrodes used as electrosurgical accessories... on . FDA classified it as Class II and gives the reason as "Encision is performing a field correction to their AEM Disposable Electrodes due to two consumer complaints reporting unanticipated alarms from the AEM monitor while using the device. This issue could result delays or...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1951-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Encision, Inc.
Product
AEM Disposable Electrodes, Device, Electrosurgical, Cutting and Coagulation and Accessories. AEM Disposable Electrodes used as electrosurgical accessories intended for ablation, removal, resection and coagulation of soft tissue in open, endoscopic and laparoscopic surgical procedures channeling monopolar high-frequency electrical current from compatible electrosurgical generators. The device is model FC03XX series. AEM Disposable Electrodes used as electrosurgical accessories intended for ablation, removal, resection and coagulation of soft tissue in open, endoscopic and laparoscopic surgical procedures channeling monopolar high-frequency electrical current from compatible electrosurgical generators.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Encision is performing a field correction to their AEM Disposable Electrodes due to two consumer complaints reporting unanticipated alarms from the AEM monitor while using the device. This issue could result delays or complications during surgical procedures.
Root cause tag
design
FDA root cause
Device Design
Action
Encision sent an Urgent Medical Device Recall letter dated June 13, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to lubricate the contact on the instrument cord with a conductive gel (to be supplied by Encision) prior to making the connection of the cord to the instrument. Instructions for applying the gel were attached. This corrective action was temporary; allowing customers to use their inventory of devices and for new product in the short term, until a design correction can resolve the problem. Customers would be contacted by their Encision Sales Representative to review this information for their specific situation and train their staff on its correction. For questions customers should call 303-339-6917 or Customer Service at 800-998-0986. For questions regarding this recall call 303-339-6917.
Quantity in commerce
48 boxes
Distribution
Worldwide Distribution - USA including CA, TX, GA, NY, NC, OH), and New Zealand.
Product codes and lots
RDB, RJB, RJE, RJH, RLG, RLI, RLJ, RLK, RLO, RLP, RLR, RLQ, SBF, SBJ.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1951-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1951-2012Class IITerminatedVoluntary: Firm initiated