Encision, Inc.

Boulder, CO, US

ENCISION, INC. appears in FDA device records in Boulder, CO. FDA records list 10 510(k) clearances between and , no PMA records and one registered establishment. As of , FDA records show 26 recalls for this company, and at least one establishment registration is current.

FDA records held for this company

510(k) clearances
10
PMA records
0
Recall records
26
Registered establishments
1

510(k) clearance history

FDA records hold 10 510(k) clearances for Encision, Inc. between 2007 and 2020. The busiest year on record is 2009, with 2. The most recent is 2020, with 1.

Clearances by year, as a table
510(k) clearance decisions for Encision, Inc., by decision year
YearClearances
20071
20092
20102
20121
20132
20141
20201
510(k) clearances on record, most recent first
K numberDeviceProduct codeDecisionDecision date
K201018Encision AEM enTouch 2X ScissorGEISubstantially Equivalent
K140006ENDOSHIELD BURN PROTECTION SYSTEMGEISubstantially Equivalent
K131949AEM DISPOSABLE SUCTION IRRIGATION ELECTRODESGEISubstantially Equivalent
K122383EM3 AEM MONITOR AEM CONNECTORSGEISubstantially Equivalent
K122580AEM DISPOSABLE SUCTION IRRIGATION HANDSET MODEL ES1700; AEM DISPOSABLE SUCTION IRRIGATION ELECTRODES MODEL ES077X SERIESGEISubstantially Equivalent
K093622AEM MONITORING SYSTEM, MODELS EM2, EMR, ES4007, ES4107, ES9005 AND ES9015GEISubstantially Equivalent
K100711SUCTION IRRIGATION ELECTRODES AND ADAPTERSGEISubstantially Equivalent
K091074AEM DISPOSABLE ELECTRODES,MODEL ESO300 SERIESGEISubstantially Equivalent
K090979DISPOSABLE HANDLE ASSEMBLES, MODELS DH0100 AND DH0500 SERIESGCJSubstantially Equivalent
K072789DISPOSABLE AEM FIXED TIP ELECTRODES AND HANDPIECESGEISubstantially Equivalent

Registered establishments

FDA registered establishments for Encision, Inc., with the product codes listed at each
RegistrationFEIEstablishmentAddressRegistration expiresProduct codes listed
17220401000117361ENCISION, INC.6797 WINCHESTER CIR., BOULDER, CO 80301 US2026

Registration numbers link to the FDA registration and listing record for that establishment.

FDA product codes

Recall records

Showing 25 of 26
Device recall records naming this company, most recent first
RecallProductReasonClassificationInitiatedStatus
Z-1951-2012AEM Disposable Electrodes, Device, Electrosurgical, Cutting and Coagulation and Accessories. AEM Disposable Electrodes used as electrosurgical accessories intended for ablation, removal, resection and coagulation of soft tissue in open, endoscopic and laparoscopic surgical procedures channeling monopolar high-frequency electrical current from compatible electrosurgical generators. The device is model FC03XX series. AEM Disposable Electrodes used as electrosurgical accessories intended for ablation, removal, resection and coagulation of soft tissue in open, endoscopic and laparoscopic surgical procedures channeling monopolar high-frequency electrical current from compatible electrosurgical generators.Encision is performing a field correction to their AEM Disposable Electrodes due to two consumer complaints reporting unanticipated alarms from the AEM monitor while using the device. This issue could result delays or complications during surgical procedures.Class IITerminated
Z-3097-2011Encision AEM Suction Irrigation Electrode, Needle Tip, REF ES3885, Encision Inc., Boulder, CO 80301.Electrode tips have been reported breaking off during use or cleaning.Terminated
Z-3096-2011Encision AEM Suction Irrigation Electrode, J-Hook, REF ES3884. Encision Inc., Boulder, CO 80301. The Suction Irrigation Electrodes and Adapters combine AEM shielding technology with suction and irrigation of fluids during laparoscopic electrosurgery to cut and coagulate tissue. The electrodes connect to a compatible electrosurgical generator via an adapter on the Encision AEM Monitor.Electrode tips have been reported breaking off during use or cleaning.Terminated
Z-3095-2011Encision AEM Suction Irrigation Electrode, L-Hook, REF ES3883, Encision Inc., Boulder, CO 80301. The Suction Irrigation Electrodes and Adapters combine AEM shielding technology with suction and irrigation of fluids during laparoscopic electrosurgery to cut and coagulate tissue. The electrodes connect to a compatible electrosurgical generator via an adapter on the Encision AEM Monitor.Electrode tips have been reported breaking off during use or cleaning.Terminated
Z-3094-2011Encision AEM Suction Irrigation Electrode, Spatula, REF ES3882, Encision Inc., Boulder, CO 80301. The Suction Irrigation Electrodes and Adapters combine AEM shielding technology with suction and irrigation of fluids during laparoscopic electrosurgery to cut and coagulate tissue. The electrodes connect to a compatible electrosurgical generator via an adapter on the Encision AEM Monitor.Electrode tips have been reported breaking off during use or cleaning.Terminated
Z-3093-2011Encision AEM Suction Irrigation Electrode, L-Diamond, REF ES3881. Encision Inc., Boulder, CO 80301. The Suction Irrigation Electrodes and Adapters combine AEM shielding technology with suction and irrigation of fluids during laparoscopic electrosurgery to cut and coagulate tissue. The electrodes connect to a compatible electrosurgical generator via an adapter on the Encision AEM Monitor.Electrode tips have been reported breaking off during use or cleaning.Terminated
Z-2400-2011ES0102-45, AEM Disposable Curved Scissors, 3/4", 45cm, Sterile R, Encision Inc., Boulder, CO 80301 Cutting tissue during electrosurgical proceduresSterility may be compromised due to holes in packaging.Terminated
Z-2399-2011ES0101-45, AEM Disposable Curved Scissors, 1/2", 45cm, Sterile R, Encision Inc., Boulder, CO 80301 Cutting tissue during electrosurgical proceduresSterility may be compromised due to holes in packaging.Terminated
Z-2398-2011ES0120, AEM Disposable Low Profile Scissors, 3/4", 35cm, Sterile R, Encision Inc., Boulder, CO 80301 Cutting tissue during electrosurgical proceduresSterility may be compromised due to holes in packaging.Terminated
Z-2397-2011ES0110, AEM Disposable Hook Scissors, 35cm, Sterile R, Encision Inc., Boulder, CO 80301 Cutting tissue during electrosurgical proceduresSterility may be compromised due to holes in packaging.Terminated
Z-2396-2011ES0102, AEM Disposable Curved Scissors, 3/4", 35cm, Sterile R, Encision Inc., Boulder, CO 80301 Cutting tissue during electrosurgical proceduresSterility may be compromised due to holes in packaging.Terminated
Z-2395-2011ES0101, AEM Disposable Curved Scissors, 1/2", 35cm, Sterile R, Encision Inc., Boulder, CO 80301 Cutting tissue during electrosurgical proceduresSterility may be compromised due to holes in packaging.Terminated
Z-1546-2010AEM Disposable Electrode, L-Wedge, 45cm, REF ES0312-45, Qty 10 each Sterile R, Manufactured for: Encision Inc., Boulder, CO 80301, made in Taiwan. Electrosurgical accessories intended, by use of monopolar high-frequency electrical current from compatible electrosurgical generators, for ablation, removal, resection and coagulation of soft tissue where associated hemostasis is required in open, endoscopic and laparoscopic surgical procedures.Electrosurgical electrode tip insulation may crack and break off during use.Terminated
Z-1545-2010AEM Disposable Electrode, L-Wedge, 35cm, REF ES0312, Qty 10 each Sterile R, Manufactured for: Encision Inc., Boulder, CO 80301, made in Taiwan. Electrosurgical accessories intended, by use of monopolar high-frequency electrical current from compatible electrosurgical generators, for ablation, removal, resection and coagulation of soft tissue where associated hemostasis is required in open, endoscopic and laparoscopic surgical procedures.Electrosurgical electrode tip insulation may crack and break off during use.Terminated
Z-1544-2010AEM Disposable Electrode, L-Diamond, 35cm, REF ES0311, Qty 10 each Sterile R, Manufactured for: Encision Inc., Boulder, CO 80301, made in Taiwan. Electrosurgical accessories intended, by use of monopolar high-frequency electrical current from compatible electrosurgical generators, for ablation, removal, resection and coagulation of soft tissue where associated hemostasis is required in open, endoscopic and laparoscopic surgical procedures.Electrosurgical electrode tip insulation may crack and break off during use.Terminated
Z-1543-2010AEM Disposable Electrode, J-Hook, 45cm, REF ES0310-45, Sterile R, Manufactured for: Encision Inc., Boulder, CO 80301, made in Taiwan. Electrosurgical accessories intended, by use of monopolar high-frequency electrical current from compatible electrosurgical generators, for ablation, removal, resection and coagulation of soft tissue where associated hemostasis is required in open, endoscopic and laparoscopic surgical procedures.Electrosurgical electrode tip insulation may crack and break off during use.Terminated
Z-1542-2010AEM Disposable Electrode, J-Hook, 35cm, REF ES0310, Qty 10 each Sterile R, Manufactured for: Encision Inc., Boulder, CO 80301, made in Taiwan. Electrosurgical accessories intended, by use of monopolar high-frequency electrical current from compatible electrosurgical generators, for ablation, removal, resection and coagulation of soft tissue where associated hemostasis is required in open, endoscopic and laparoscopic surgical procedures.Electrosurgical electrode tip insulation may crack and break off during use.Terminated
Z-1541-2010AEM Disposable Electrode, Flat J-Hook, 35cm, REF ES0309, Qty 10 each Sterile R, Manufactured for: Encision Inc., Boulder, CO 80301, made in Taiwan. Electrosurgical accessories intended, by use of monopolar high-frequency electrical current from compatible electrosurgical generators, for ablation, removal, resection and coagulation of soft tissue where associated hemostasis is required in open, endoscopic and laparoscopic surgical procedures.Electrosurgical electrode tip insulation may crack and break off during use.Terminated
Z-1540-2010AEM Disposable Electrode, Spatula, 45cm, REF ES0301-45, Qty 10 each Sterile R, Manufactured for: Encision Inc., Boulder, CO 80301, made in Taiwan. Electrosurgical accessories intended, by use of monopolar high-frequency electrical current from compatible electrosurgical generators, for ablation, removal, resection and coagulation of soft tissue where associated hemostasis is required in open, endoscopic and laparoscopic surgical procedures.Electrosurgical electrode tip insulation may crack and break off during use.Terminated
Z-1539-2010AEM Disposable Electrode, Spatula, 35cm, REF ES0301, Qty: 10 each Sterile R, Manufactured for: Encision Inc., Boulder, CO 80301, made in Taiwan. Electrosurgical accessories intended, by use of monopolar high-frequency electrical current from compatible electrosurgical generators, for ablation, removal, resection and coagulation of soft tissue where associated hemostasis is required in open, endoscopic and laparoscopic surgical procedures.Electrosurgical electrode tip insulation may crack and break off during use.Terminated
Z-1132-04Encision ES4107 Disposable AEM Cord, Sterile.Firm found pinholes in packaging which would compromise sterility.Terminated
Z-1074-04AEM Fixed-Tip L-Diamond Electrode, 5mm, Reusable.Electrosurgical tip may break or come off during use.Terminated
Z-1052-03EM2M NC AEM Monitor, (NON-COM)Use of NC versions of AEM monitor presents burn hazard when used with single pad and capacitively coupled patient return electrodes.Terminated
Z-1051-03EM2+A NC AEM Monotor (NON-COM)Use of NC versions of AEM monitor presents burn hazard when used with single pad and capacitively coupled patient return electrodes.Terminated
Z-1050-03EM2 NC AEM Monitor (NON-COM)Use of NC versions of AEM monitor presents burn hazard when used with single pad and capacitively coupled patient return electrodes.Terminated

Verification record

There is no verification record for this company. Most companies on this site have none: a record exists only where a company asked to be verified and MDM ran the checks, so the absence of one says nothing about the company.

Where there is a record, this is where it appears: what was checked, when it was checked, and the current status. An MDM Verified badge on another website links to a verify address on this domain that resolves to this section. A badge that does not resolve here is not ours.

What MDM checks, and what it does not

MDM coverage

MDM coverage of this company lists here, newest first. Submit news.

Is this your company?

This page is built from public FDA records. Company pages are free and factual, and stay free whether or not anyone claims them. Membership never changes editorial coverage or vetting outcomes.

Claiming lets Encision, Inc. edit what it says about itself and list the categories it works in. It grants no badge: claiming is not verification.

Claim this page

Data as of , FDA export datePage updated Source: openFDA, public domain