Z-2278-2012Class II

EXCELART Vantage, Atlas, and Titan MRI Systems, MRT-1503 and MRT-1504. Imaging of the Whole Body, Fluid...

Canon Medical Systems USA, Inc. / initiated 2008-05-13 / Terminated / root cause: manufacturing process

Toshiba American Medical Systems Inc began a recall of EXCELART Vantage, Atlas, and Titan MRI Systems, MRT-1503 and MRT-1504. Imaging of the Whole Body, Fluid Visualization, 2D/3D Imaging, MR Angiography/MR... on . FDA classified it as Class II and gives the reason as "The status display in the sequence queue window may not change from "Current" to "Done"". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2278-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
EXCELART Vantage, Atlas, and Titan MRI Systems, MRT-1503 and MRT-1504. Imaging of the Whole Body, Fluid Visualization, 2D/3D Imaging, MR Angiography/MR Vascular Imaging, Blood Oxygenation Level Dependent Imaging, Perfusion I Diffusion Imaging, and Proton Spectroscopy.
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
The status display in the sequence queue window may not change from "Current" to "Done".
Root cause tag
manufacturing process
FDA root cause
Process design
Action
A recall notification letter titled: "URGENT: MEDICAL DEVICE CORRECTION," was sent to customers. The letter stated the problem, corrective action, and (instructions) request to customers. Customers were instructed to perform the temporary recovery procedures if the problem occurs. Additionally, the acknowledgement sheet should be completed and faxed back. The information should be shared with all users, including the bio-medical or clinical engineering departments. Questions pertaining to the information for this recall were to be directed to (800) 421-1968.
Quantity in commerce
54
Distribution
Nationwide Distribution
Product codes and lots
Models MRT-1503 and MRT-1504  " MRT-1503 systems with software V9.01 *R240 through V9.20*R242 " MRT-1504 systems with software V9.21 *R24

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2278-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2278-2012Class IITerminatedVoluntary: Firm initiated