Toshiba Ultra-Short Magnetic Resonance Imaging System, EXCELART Vantage. MRT-1503/P5-Vantage MRI System. MR...
Toshiba American Medical Systems Inc began a recall of Toshiba Ultra-Short Magnetic Resonance Imaging System, EXCELART Vantage. MRT-1503/P5-Vantage MRI System. MR imaging on . FDA classified it as Class II and gives the reason as "The firm recalled due to a potentail problem with the position of the hold in the lug teminals on the power cables connected to the terminal blocks on the gradient coil". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2315-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- Toshiba Ultra-Short Magnetic Resonance Imaging System, EXCELART Vantage. MRT-1503/P5-Vantage MRI System. MR imaging.
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- The firm recalled due to a potentail problem with the position of the hold in the lug teminals on the power cables connected to the terminal blocks on the gradient coil.
- FDA root cause
- Component change control
- Action
- Toshiba America Medical Systems, Inc, sent a Urgent Medical Device Correction letter dated February 21, 2012, via USPS return receipt mail to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed if you subsequently find that images are abnormally noisy compared to images acquired under the same conditions in the system before this modification was performed, contact your Toshiba Service Representative. Please share this information with all users and reviewing radiologist as well as clinical engineering or biomedical group at your facility. Please complete and return the attached form and fax it to the toll free number at the top of the form. This form can also be sent via email [email protected]. Contact information was also provided (800) 421-1968 or contact your local Toshiba Representative at (800) 521-1968.
- Quantity in commerce
- 3 units
- Distribution
- US (nationwide) distribution including the states of Ohio and Texas.
- Product codes and lots
- MRT-1503/P5-Vantage MRI System. Serial Codes: P5A0632002, P5B06Y2005, P5B0712006, P3F11Z2143.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2315-2012 | Class II | Terminated | Voluntary: Firm initiated |