TSX-301A/2 Aquilion One CT System. The TSX-30IAl2 is a whole body CT scanner. This device captures cross...
Toshiba American Medical Systems Inc began a recall of TSX-301A/2 Aquilion One CT System. The TSX-30IAl2 is a whole body CT scanner. This device captures cross sectional volume data sets. The device consists of a... on . FDA classified it as Class II and gives the reason as "The SureStart function may not operate in the usual manner, and you may obtain abnormal analysis results when using the cerebral blood flow analysis function". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2468-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- TSX-301A/2 Aquilion One CT System. The TSX-30IAl2 is a whole body CT scanner. This device captures cross sectional volume data sets. The device consists of a gantry, patient couch (table) and peripheral cabinets used for data processing and display. This device is intended to produce cross sectional volume sets of the anatomy. Additionally the system, when used by a trained professional, is capable of proving data for dynamic and perfusion studies of organs and extremities.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- The SureStart function may not operate in the usual manner, and you may obtain abnormal analysis results when using the cerebral blood flow analysis function.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Toshiba American Medical Systems, Inc. sent an "URGENT: MEDICAL DEVICE CORRECTION" letter to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. customers: The firm advised customers to take the following temporary measures to avoid issues until a corrective action has been taken: I. In SureStart: " Perform scanning with Observation Direction set to VFF when the Patient Insertion Direction and Observation direction are FF. " Perform scanning with the Observation direction set to VFH when the Patient Insertion Direction and Observation Direction are HF. 2. When acquiring images for 4D perfusion analysis: " Perform scanning with the Scan Direction set to OUT when the Patient Insertion Direction and Scan Direction are HF. Perform scanning with the scan direction set to IN when the Patient Insertion Direction and Scan Direction are FF. Contact the firm at (800) 421-1968 for questions regarding this issue.
- Quantity in commerce
- 3 units
- Distribution
- Nationwide distribution: USA including states of: MA, MD, and NV.
- Product codes and lots
- Item number TSX-301 A/2D Serial Numbers: 2DA0792004, 2DA07X2005, 2DA07Y2010, 2DA0792002, 2DA07X2008, 2DA0842023.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2468-2012 | Class II | Terminated | Voluntary: Firm initiated |