Z-2468-2012Class II

TSX-301A/2 Aquilion One CT System. The TSX-30IAl2 is a whole body CT scanner. This device captures cross...

Canon Medical Systems USA, Inc. / initiated 2008-05-13 / Terminated / root cause: software

Toshiba American Medical Systems Inc began a recall of TSX-301A/2 Aquilion One CT System. The TSX-30IAl2 is a whole body CT scanner. This device captures cross sectional volume data sets. The device consists of a... on . FDA classified it as Class II and gives the reason as "The SureStart function may not operate in the usual manner, and you may obtain abnormal analysis results when using the cerebral blood flow analysis function". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2468-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
TSX-301A/2 Aquilion One CT System. The TSX-30IAl2 is a whole body CT scanner. This device captures cross sectional volume data sets. The device consists of a gantry, patient couch (table) and peripheral cabinets used for data processing and display. This device is intended to produce cross sectional volume sets of the anatomy. Additionally the system, when used by a trained professional, is capable of proving data for dynamic and perfusion studies of organs and extremities.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
The SureStart function may not operate in the usual manner, and you may obtain abnormal analysis results when using the cerebral blood flow analysis function.
Root cause tag
software
FDA root cause
Software design
Action
Toshiba American Medical Systems, Inc. sent an "URGENT: MEDICAL DEVICE CORRECTION" letter to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. customers: The firm advised customers to take the following temporary measures to avoid issues until a corrective action has been taken: I. In SureStart: " Perform scanning with Observation Direction set to VFF when the Patient Insertion Direction and Observation direction are FF. " Perform scanning with the Observation direction set to VFH when the Patient Insertion Direction and Observation Direction are HF. 2. When acquiring images for 4D perfusion analysis: " Perform scanning with the Scan Direction set to OUT when the Patient Insertion Direction and Scan Direction are HF. Perform scanning with the scan direction set to IN when the Patient Insertion Direction and Scan Direction are FF. Contact the firm at (800) 421-1968 for questions regarding this issue.
Quantity in commerce
3 units
Distribution
Nationwide distribution: USA including states of: MA, MD, and NV.
Product codes and lots
Item number TSX-301 A/2D Serial Numbers: 2DA0792004, 2DA07X2005, 2DA07Y2010, 2DA0792002, 2DA07X2008, 2DA0842023.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2468-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2468-2012Class IITerminatedVoluntary: Firm initiated