Z-1101-2013Class II

MRT-1504/A4 - Titan X MRI System MRT-30 I0A/5 - Titan 3T MRI System

Canon Medical Systems USA, Inc. / initiated 2012-10-15 / Terminated / root cause: manufacturing process

Toshiba American Medical Systems Inc began a recall of MRT-1504/A4 - Titan X MRI System MRT-30 I0A/5 - Titan 3T MRI System Product Usage: MRI System on . FDA classified it as Class II and gives the reason as "It was found that there is a problem with the LCD monitor used in the console. The LCD monitor may become blank during use. Specifically, the backlight of the LCD monitor may turn off while the system is in use, causing...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1101-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
MRT-1504/A4 - Titan X MRI System MRT-30 I0A/5 - Titan 3T MRI System Product Usage: MRI System.
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
It was found that there is a problem with the LCD monitor used in the console. The LCD monitor may become blank during use. Specifically, the backlight of the LCD monitor may turn off while the system is in use, causing the LCD monitor to become blank. However, the other parts of the system continue to operate normally, and the risk of a serious health hazard is therefore very low. The LCD Moni
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Toshiba sent an Urgent Medical Device Correction letter dated October 1, 2012 via USPS return receipt mail to all affected customers . The letter identified the affected product, problem and actions to be taken. The letter contains a return reply form that is to be faxed to TAMS for retention. For questions contact your local Toshiba representative at 800-521-1968.
Quantity in commerce
27 units
Distribution
USA Nationwide Distribution including the states of : IL, NH, MA, MI, MO, PR, NV, NM, MN, ID, NY, OH, NC, AZ, MS, LA, OR, PA, CA, and CT.
Product codes and lots
Product Numbers: MRT-1504/A4, MRT-1504/S4, MRT-3010/A5, MZKT-UPKHE/S1. Serial Numbers unique to each consignee.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1101-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1101-2013Class IITerminatedVoluntary: Firm initiated