MRT-1504/A4 - Titan X MRI System MRT-30 I0A/5 - Titan 3T MRI System
Toshiba American Medical Systems Inc began a recall of MRT-1504/A4 - Titan X MRI System MRT-30 I0A/5 - Titan 3T MRI System Product Usage: MRI System on . FDA classified it as Class II and gives the reason as "It was found that there is a problem with the LCD monitor used in the console. The LCD monitor may become blank during use. Specifically, the backlight of the LCD monitor may turn off while the system is in use, causing...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1101-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- MRT-1504/A4 - Titan X MRI System MRT-30 I0A/5 - Titan 3T MRI System Product Usage: MRI System.
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- It was found that there is a problem with the LCD monitor used in the console. The LCD monitor may become blank during use. Specifically, the backlight of the LCD monitor may turn off while the system is in use, causing the LCD monitor to become blank. However, the other parts of the system continue to operate normally, and the risk of a serious health hazard is therefore very low. The LCD Moni
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Toshiba sent an Urgent Medical Device Correction letter dated October 1, 2012 via USPS return receipt mail to all affected customers . The letter identified the affected product, problem and actions to be taken. The letter contains a return reply form that is to be faxed to TAMS for retention. For questions contact your local Toshiba representative at 800-521-1968.
- Quantity in commerce
- 27 units
- Distribution
- USA Nationwide Distribution including the states of : IL, NH, MA, MI, MO, PR, NV, NM, MN, ID, NY, OH, NC, AZ, MS, LA, OR, PA, CA, and CT.
- Product codes and lots
- Product Numbers: MRT-1504/A4, MRT-1504/S4, MRT-3010/A5, MZKT-UPKHE/S1. Serial Numbers unique to each consignee.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1101-2013 | Class II | Terminated | Voluntary: Firm initiated |