Z-0201-2014Class II

Titan 3T MRI System, MRT-3010A/5

Canon Medical Systems USA, Inc. / initiated 2013-10-28 / Terminated / root cause: manufacturing process

Toshiba American Medical Systems Inc began a recall of Titan 3T MRI System, MRT-3010A/5. Product Usage: Usage: MRI system on . FDA classified it as Class II and gives the reason as "Toshiba America Medical Systems, Inc. is recalling due to a potential problem with Specific Absorption Rate (SAR) values when using the Toshiba Titan 3T System. When the system is used in combination with either the QD...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0201-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
Titan 3T MRI System, MRT-3010A/5. Product Usage: Usage: MRI system.
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
Toshiba America Medical Systems, Inc. is recalling due to a potential problem with Specific Absorption Rate (SAR) values when using the Toshiba Titan 3T System. When the system is used in combination with either the QD Head Coil (MJQH-142A) or the Large Knee SPEEDER (MJAJ-182A), SAR values lower than the actual SAR values may be displayed on the operating monitor.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Toshiba America Medical Systems, Inc. sent an "URGENT: MEDICAL DEVICE CORRECTION". letter dated October 28, 2013 via USPS return receipt mail. The letter contains a return reply form that is to be faxed to TAMS for retention. The letter listed Device, Serial Number, reason for recall, problem, corrective actions, request to customers, and contact information.
Quantity in commerce
18 units
Distribution
USA Nationwide Distribution in the states of: CA, FL, GA, IA, NV, NY, LA, IN, PA, OR, TX, and WI.
Product codes and lots
Lot numbers not utilized. Serial Numbers:A5B1192003 A5B1192004, A5B11Y2005, A5B11Z2007, A5C1212008, A5C1212009, A5C1222010, A5C1232011, A5C1252012, A5C1262013, A5C1282014, A5C1292015, A5C12X2016, A5C12Y2017, A5C12Z2018, A5D1312019, A5D1322020, A5D1362021, XIB-A5A1082001, XIB-A5B11Y2006, A5D1372022, A5B1182002.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0201-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0201-2014Class IITerminatedVoluntary: Firm initiated