Titan 3T MRI System, MRT-3010A/5
Toshiba American Medical Systems Inc began a recall of Titan 3T MRI System, MRT-3010A/5. Product Usage: Usage: MRI system on . FDA classified it as Class II and gives the reason as "Toshiba America Medical Systems, Inc. is recalling due to a potential problem with Specific Absorption Rate (SAR) values when using the Toshiba Titan 3T System. When the system is used in combination with either the QD...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0201-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- Titan 3T MRI System, MRT-3010A/5. Product Usage: Usage: MRI system.
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- Toshiba America Medical Systems, Inc. is recalling due to a potential problem with Specific Absorption Rate (SAR) values when using the Toshiba Titan 3T System. When the system is used in combination with either the QD Head Coil (MJQH-142A) or the Large Knee SPEEDER (MJAJ-182A), SAR values lower than the actual SAR values may be displayed on the operating monitor.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Toshiba America Medical Systems, Inc. sent an "URGENT: MEDICAL DEVICE CORRECTION". letter dated October 28, 2013 via USPS return receipt mail. The letter contains a return reply form that is to be faxed to TAMS for retention. The letter listed Device, Serial Number, reason for recall, problem, corrective actions, request to customers, and contact information.
- Quantity in commerce
- 18 units
- Distribution
- USA Nationwide Distribution in the states of: CA, FL, GA, IA, NV, NY, LA, IN, PA, OR, TX, and WI.
- Product codes and lots
- Lot numbers not utilized. Serial Numbers:A5B1192003 A5B1192004, A5B11Y2005, A5B11Z2007, A5C1212008, A5C1212009, A5C1222010, A5C1232011, A5C1252012, A5C1262013, A5C1282014, A5C1292015, A5C12X2016, A5C12Y2017, A5C12Z2018, A5D1312019, A5D1322020, A5D1362021, XIB-A5A1082001, XIB-A5B11Y2006, A5D1372022, A5B1182002.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0201-2014 | Class II | Terminated | Voluntary: Firm initiated |