Z-0908-2014Class II

Product Name: Model 7305- Mosaic Aortic Obturator The Mosaic Aortic Obturators are used with the Mosaic...

Medtronic Heart Valves Division / initiated 2014-01-09 / Terminated / root cause: design

Medtronic Cardiovascular Surgery-the Heart Valve Division began a recall of Product Name: Model 7305- Mosaic Aortic Obturator The Mosaic Aortic Obturators are used with the Mosaic Porcine Aortic Bioprostheses on . FDA classified it as Class II and gives the reason as "Medtronic, Inc. initiated a recall of Medtronic Mosaic Aortic Obturator Model 7305 used with the Mosaic Porcine Aortic Bioprosthesis, Model Number 305, because of higher-than-expected transvalvular gradients occurring...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0908-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Heart Valves Division
Product
Product Name: Model 7305- Mosaic Aortic Obturator The Mosaic Aortic Obturators are used with the Mosaic Porcine Aortic Bioprostheses.
Product code
LWR Heart-Valve, Non-Allograft Tissue
Reason for recall
Medtronic, Inc. initiated a recall of Medtronic Mosaic Aortic Obturator Model 7305 used with the Mosaic Porcine Aortic Bioprosthesis, Model Number 305, because of higher-than-expected transvalvular gradients occurring post implant.
Root cause tag
design
FDA root cause
Device Design
Action
Medtronic sent an Customer Notification letter dated January 9, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. Medtronic recommends surgeons continue to remain diligent in selecting the appropriate size of Mosaic Aortic Bioprosthesis. Adoption of the updated Mosaic Aortic Bioprosthesis sizing chart (iEOA chart) and Mosaic Obturators/Sizers will facilitate optimal aortic valve size selection. Your Medtronic Field Representative will be contacting you about receiving a new Mosaic Aortic Obturator/Sizer set and sizing chart. If you would prefer to order the new Mosaic Aortic Obturator/Sizer set directly through Medtronic Customer Service, contact them at 1-800-848-9300. Medtronic will notify all applicable regulatory agencies, as required, about this matter. Please share this notification with others in your organization as appropriate. We appreciate your review of this notification and apologize for the inconvenience that it may cause. If you have any questions, please contact your Medtronic Field Representative or Lifeline Technical Services at 1-877-526-7890.
Quantity in commerce
8276 units
Distribution
Worldwide Distribution - USA (nationwide)
Product codes and lots
Model 7305.  Lot/Serial Numbers:  The catalog numbers for the Obturator/Sizer are 7305C, 7305UX, and 7305OD.  Scope of this notification includes all Mosaic Obturators/Sizers, used with Mosaic Porcine Aortic Bioprosthesis implants.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0908-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0908-2014Class IITerminatedVoluntary: Firm initiated