Medtronic Heart Valves Division
Santa Ana, CA, US
Medtronic Heart Valves Division appears in FDA device records in Santa Ana, CA. FDA records list no 510(k) clearances, no PMA records and one registered establishment. As of , FDA records show 13 recalls for this company, and at least one establishment registration is current.
FDA records held for this company
FDA records under these names
2 namesFDA records resolved to Medtronic Heart Valves Division carry these names as well, as FDA wrote them. Each is the firm named on a registration, a 510(k), a PMA, a recall or an enforcement report that was joined to this company on a shared FDA identifier or on a matching name.
- Medtronic Cardiovascular Surgery-the Heart Valve Divisionenforcement report, recall
- Medtronic CoreValve LLCenforcement report, recall
510(k) clearance history
No 510(k) clearance records for this company in the ingested FDA data.
No 510(k) clearance records in the ingested FDA data.
Registered establishments
| Registration | FEI | Establishment | Address | Registration expires | Product codes listed |
|---|---|---|---|---|---|
| 2025587 | 2025587 | Medtronic Heart Valves Division | 1851 EAST DEERE AVE., Santa Ana, CA 92705 US | 2026 | and 2 more, listed in full below |
Registration numbers link to the FDA registration and listing record for that establishment.
FDA product codes
- NPTAortic Valve, Prosthesis, Percutaneously Delivered
- LDFElectrode, Pacemaker, Temporary
- LWQHeart-Valve, Mechanical
- LWRHeart-Valve, Non-Allograft Tissue
- DTJHolder, Heart-Valve, Prosthesis
- DWSInstruments, Surgical, Cardiovascular
- NPVPulmonary Valve Prosthesis Percutaneously Delivered
- MWHPulmonic Valved Conduit
- DYEReplacement Heart-Valve
- KRHRing, Annuloplasty
- DTISizer, Heart-Valve, Prosthesis
- DQXWire, Guide, Catheter
Recall records
| Recall | Product | Reason | Classification | Initiated | Status |
|---|---|---|---|---|---|
| Z-2624-2026 | Medtronic Harmony Delivery Catheter System | Devices containing inner shaft assemblies from specific manufacturing lots may have an increased potential for distal tip detachment from the delivery system under certain procedural or anatomical conditions due to a manufacturing issue. | Class I | Open, Classified | |
| Z-2043-2021 | CoreValve Evolut PRO PLUS Delivery Catheter System, REF: D-EVPROP2329US; D-EVPROP34US; | Due to reports of actuator separation for various delivery catheter systems. Affected products were manufactured before July 11, 2020. | Class II | Open, Classified | |
| Z-2042-2021 | CoreValve Evolut PRO Delivery Catheter System, REF: ENVPRO-14; ENVPRO-16; | Due to reports of actuator separation for various delivery catheter systems. Affected products were manufactured before July 11, 2020. | Class II | Open, Classified | |
| Z-2041-2021 | CoreValve Evolut R Delivery Catheter System, REF: ENVEOR-L; ENVEOR-N; | Due to reports of actuator separation for various delivery catheter systems. Affected products were manufactured before July 11, 2020. | Class II | Open, Classified | |
| Z-0561-2021 | Evolut (TM) PRO+ System. With the following model numbers: a. EVPROPLUS-23US, Labeled as VLV EVPROPLUS-23US COMM JPN, VLV EVPROPLUS-23 COMM US; b. EVPROPLUS-26US, Labeled as VLV EVPROPLUS-26US COMM JPN, VLV EVPROPLUS-26 COMM US; c. EVPROPLUS-29US, Labeled as VLV EVPROPLUS-29US COMM JPN, VLV EVPROPLUS-29 COMM US; d. EVPROPLUS-34US, Labeled as VLV EVPROPLUS-34 COMM US, VLV EVPROPLUS-34 COMM US SA. | Potential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatation (PID). | Class II | Terminated | |
| Z-0560-2021 | Evolut (TM) PRO System. With the following model numbers: a. EVOLUTPRO-23, Labeled as: VLV EVOLUTPRO-23 PRO COMM 29L OUS, VLV EVOLUTPRO-23 BLUE 29L COMM OUS, VLV EVOLUTPRO-23 BLUE COMM JPN, VLV EVOLUTPRO-23 BLUE 6L COMM OUS, VLV EVOLUTPRO-23 PRO COMM 29L CAN, VLV EVOLUTPRO-23 PRO COMM 29L OUS SA, VLV EVOLUTPRO-23 PRO COMM 29L AUS; b. EVOLUTPRO-23-US, Labeled as: VLV EVOLUTPRO-23-US BLUE COMM US, VLV EVOLUTPRO-23-US PRO BLNK US; c. EVOLUTPRO-26, Labeled as: VLV EVOLUTPRO-26 PRO COMM 29L OUS, VLV EVOLUTPRO-26 BLUE 6L COMM OUS, VLV EVOLUTPRO-26 PRO COMM 29L AUS, VLV EVOLUTPRO-26 BLUE COMM JPN, VLV EVOLUTPRO-26 BLUE 29L COMM OUS, VLV EVOLUTPRO-26 PRO COMM 29L CAN, VLV EVOLUTPRO-26 PRO COMM 29L OUS SA; d. EVOLUTPRO-26-US, Labeled as: VLV EVOLUTPRO-26-US BLUE COMM US, VLV EVOLUTPRO-26-US PRO BLNK US; e. EVOLUTPRO-29, Labeled as: VLV EVOLUTPRO-29 PRO COMM 29L OUS, VLV EVOLUTPRO-29 BLUE 29L COMM OUS, VLV EVOLUTPRO-29 BLUE 6L COMM OUS, VLV EVOLUTPRO-29 PRO COMM 29L AUS, VLV EVOLUTPRO-29 BLUE COMM JPN, VLV EVOLUTPRO-29 PRO COMM 29L CAN, VLV EVOLUTPRO-29 PRO COMM 29L OUS SA; f. EVOLUTPRO-29-US, Labeled as: VLV EVOLUTPRO-29-US BLUE COMM US, VLV EVOLUTPRO-29-US PRO BLNK US g. , Labeled as: | Potential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatation (PID). | Class II | Terminated | |
| Z-0559-2021 | Evolut (TM) R System. With the following model numbers: a. EVOLUTR-23, Labeled as VLV EVOLUTR-23 VT IR 6L COMM OUS, VLV EVOLUTR-23 TAV EVOLUT R CE 29L MX, VLV EVOLUTR-23 TAV CE AU 2Y MX, VLV EVOLUTR-23 TAV JPN MX, VLV EVOLUTR-23 GREY 29L COMM OUS, VLV EVOLUTR-23 TAV EVOLUT R CE 7L MX, VLV EVOLUTR-23 TAV EVOLUT R CE 26L MX, VLV EVOLUTR-23 TAV CA 2Y MX, VLV EVOLUTR-23 GREY 6L COMM OUS, VLV EVOLUTR-23-US TAV 23 US COMM MX; b. EVOLUTR-23-C, Labeled as VLV EVOLUTR-23-C TAV EVOLUT R US IDE MEX; c. EVOLUTR-23-US, Labeled as VLV EVOLUTR-23-US GREY COMM US, VLV EVOLUTR-23-US TAV 23 US COMM MX; d. EVOLUTR-26, Labeled as VLV EVOLUTR-26 TAV EVOLUT R CE 29L MX VLV EVOLUTR-26 VT IR 6L COMM OUS, VLV EVOLUTR-26 TAV EVOLUT R CE 7L MX, LV EVOLUTR-26 TAV CE AU 2Y MX, VLV EVOLUTR-26 TAV JPN MX, VLV EVOLUTR-26 GREY 29L COMM OUS, VLV EVOLUTR-26-US TAV 26 US COMM MX, VLV EVOLUTR-26 TAV CLN OUS MX, VLV EVOLUTR-26 TAV CA 2Y MX, VLV EVOLUTR-26 TAV CE CA 2Y MX, VLV EVOLUTR-26 GREY 6L COMM OUS; e. EVOLUTR-26-C, Labeled as VLV EVOLUTR-26-C TAV EVOLUT R US IDE MEX, VLV EVOLUTR-26-C TAV CLN US MX; f. EVOLUTR-26-US, Labeled as VLV EVOLUTR-26 TAV EVOLUT R CE 29L MX, VLV EVOLUTR-26-US GREY COMM US, VLV EVOLUTR-26-US TAV 26 US COMM MX; g. EVOLUTR-29, Labeled as VLV EVOLUTR-29 TAV EVOLUT R CE 29L MX, VLV EVOLUTR-29 GREY 6L COMM OUS, VLV EVOLUTR-29 VT IR 6L COMM OUS, VLV EVOLUTR-29 TAV CE AU 2Y MX, VLV EVOLUTR-29 TAV JPN MX, VLV EVOLUTR-29 GREY 29L COMM OUS, VLV EVOLUTR-29 TAV EVOLUT R CE 7L MX, VLV EVOLUTR-29 TAV CLN OUS MX, VLV EVOLUTR-29 TAV CA 2Y MX, VLV EVOLUTR-29 TAV CE CA 2Y MX, VLV EVOLUTR-29-US TAV 29 US COMM MX; h. EVOLUTR-29-C, Labeled as VLV EVOLUTR-29-C TAV EVOLUT R US IDE MEX, LV EVOLUTR-29-C TAV CLN US MX, VLV EVOLUTR-29-US TAV 29 US COMM MX; i. EVOLUTR-29-US, Labeled as VLV EVOLUTR-29-US GREY COMM US, VLV EVOLUTR-29-US TAV 29 US COMM MX, VLV EVOLUTR-29 TAV EVOLUT R CE 29L MX; j. EVOLUTR-34, Labeled as VLV EVOLUTR-34 BLUE 29L COMM OUS, VLV EVOLUTR-34 VT IR 6L COMM OUS, VLV EVOLUTR-34 BLUE 6L COMM OUS, VLV EVOLUTR-34 TAV CE AU 2Y MX, VLV EVOLUTR-34 TAV JPN MX, VLV EVOLUTR-34 TAV CLN OUS MX, VLV EVOLUTR-34 TAV CA 2Y MX, VLV EVOLUTR-34 TAV CE CA 2Y MX, VLV EVOLUTR-34 TAV EVOLUT R CE 29L 1Y MX, VLV EVOLUTR-34 TAV EVO R CE 29L 2Y MX SA, DRAPE 9732722 TUBE STERILE O-ARM 20PK; k. EVOLUTR-34-C, Labeled as VLV EVOLUTR-34-C TAV EVOLUT R US IDE MX, VLV EVOLUTR-34-C TAV CLN US MX; l. EVOLUTR-34-US, Labeled as VLV EVOLUTR-34-US BLUE COMM US, VLV EVOLUTR-34-US TAV 34 US COMM MX, VLV EVOLUTR-34 TAV EVOLUT R CE 29L 1Y MX, VLV EVOLUTR-34-US TAV 34 US COMM MX SA. | Potential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatation (PID). | Class II | Terminated | |
| Z-1450-2020 | Medtronic Simulus 850 Semi Rigid Ring/Band Accessory Kit Used to size a patient's mitral or tricuspid valve in cardiac procedures. | Demonstration Simulus Reusable Sizer Accessory Kits were being incorrectly sourced to customers. The demonstration product (product used within a wet lab) could have come in contact with non-human (porcine) tissue. | Class II | Open, Classified | |
| Z-1449-2020 | Medtronic Simulus 750 Flexible Ring/Band Accessory Kit - Product Usage: Used to size a patient's mitral or tricuspid valve in cardiac procedures. | Demonstration Simulus Reusable Sizer Accessory Kits were being incorrectly sourced to customers. The demonstration product (product used within a wet lab) could have come in contact with non-human (porcine) tissue. | Class II | Open, Classified | |
| Z-2546-2016 | Medtronic EnVeo R Delivery Catheter System (DCS), Models: ENVEOR-L, ENVEOR-L-C, and ENVEOR-US. Designed to replace the native or surgical bioprosthetic aortic heart valve without open heart surgery and without concomitant surgical removal of the failed valve. | Potential for vascular trauma with the use of the Medtronic EnVeo R Delivery Catheter System (DCS). | Class II | Terminated | |
| Z-2264-2015 | EnVeo R Loading System, Model No. LS-EnVeoR-23, LS-EnVeoR-2629, LS-EnVeoR-23-C, LS-EnVeoR-2629-C. Cardiovascular valve delivery system for use with the Medtronic CoreValve Evolut R system. | Medtronic is recalling the EnVeo R Loading Systems because of the presence of particulate. | Class I | Terminated | |
| Z-2623-2014 | Hancock Apical Left Ventricle Connector, Model No. H174Axx, Catalog No. 0174ACP012. Used use in conjunction with Hancock Conduit Model 105 to provide an alternate method for relief of left ventricular hypertension. | Medtronic has decided to discontinue production and distribution of this low-volume product. | Class II | Terminated | |
| Z-0908-2014 | Product Name: Model 7305- Mosaic Aortic Obturator The Mosaic Aortic Obturators are used with the Mosaic Porcine Aortic Bioprostheses. | Medtronic, Inc. initiated a recall of Medtronic Mosaic Aortic Obturator Model 7305 used with the Mosaic Porcine Aortic Bioprosthesis, Model Number 305, because of higher-than-expected transvalvular gradients occurring post implant. | Class II | Terminated |
Verification record
There is no verification record for this company. Most companies on this site have none: a record exists only where a company asked to be verified and MDM ran the checks, so the absence of one says nothing about the company.
Where there is a record, this is where it appears: what was checked, when it was checked, and the current status. An MDM Verified badge on another website links to a verify address on this domain that resolves to this section. A badge that does not resolve here is not ours.
MDM coverage
MDM coverage of this company lists here, newest first. Submit news.
Is this your company?
This page is built from public FDA records. Company pages are free and factual, and stay free whether or not anyone claims them. Membership never changes editorial coverage or vetting outcomes.