Z-0472-2015Class II

Cardiac Function Analysis Software Product Data No. MPDCT0301EAD CSCF-003A The software is applicable to the...

Canon Medical Systems USA, Inc. / initiated 2014-11-05 / Terminated / root cause: software

Toshiba American Medical Systems Inc began a recall of Cardiac Function Analysis Software Product Data No. MPDCT0301EAD CSCF-003A The software is applicable to the following CT systems: Aquilion ONE TSX-301A/2D... on . FDA classified it as Class II and gives the reason as "Potential problem with the cardiac function analysis software (CFA). Incorrect analysis results may be displayed in the Function Parameters for the Entire Heart displayed as analysis results of the CFA and in a...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0472-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
Cardiac Function Analysis Software Product Data No. MPDCT0301EAD CSCF-003A The software is applicable to the following CT systems: Aquilion ONE TSX-301A/2D X-ray CT system with a 320-row 0.5 mm detector is provided with a 7.5-MHU large-capacity X-ray tube unit, and permits scanning over large areas at the high-speed of 0.35 s per rotation. The system features are; 160-mm-wide area detector, station, gantry and patient couch operation controls, image reconstruction, high-speed volume scan, high-quality images, selectable image slice thickness, dual-monitor system, improved image diagnosis functions, high-speed volume data workflow, exposure reduction, ECG-gated scan and reconstruction, and SUREFluoro.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Potential problem with the cardiac function analysis software (CFA). Incorrect analysis results may be displayed in the Function Parameters for the Entire Heart displayed as analysis results of the CFA and in a Left-Ventricular Volume Curve generated based on some of those parameters.
Root cause tag
software
FDA root cause
Software design
Action
Toshiba sent an Urgent Medical Device Correction letter dated November 5, 2014, to all affected customers. Customers were asked to refrain from using the function parameters for the entire heart and the left-ventricular volume curve until corrective actions have been taken. Customers were instructed to share this information with all users and reviewing radiologist as well as clinical engineering or Biomedical group at facility. Please complete and return the attached form and fax it to the toll free number at the top of the form 877-349-3054. This form can also be sent via email to [email protected]. Revised software to prevent occurrence of this problem will be installed on the system. When the new software becomes available the Toshiba service representative will contact for an appointment to install it on the system. Customers with questions were instructed to call 800-421-1968 or contact their local Toshiba Representative at 800-521-1968.
Quantity in commerce
2 units within the U.S. only
Distribution
US Nationwide Distribution to CA, MA, MD
Product codes and lots
Item CFA Software Serial Number: Description TSX-301A/2D 1BA07X2002 Aquilion One TSX-301A/2D 1BA07X2001 Aquilion One

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0472-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0471-2015Class IITerminatedVoluntary: Firm initiated
Z-0472-2015Class IITerminatedVoluntary: Firm initiated